UT3500 series
User Manual
59
0.5%rdg.
±
20dgt
High-
Speed
±
0.5%rdg.
±
40dgt
±
1%rdg.
±
8dgt
Temperature Index
(
±
0.05%rdg.
±
1dgt)/
℃
(
±
0.05%rdg.
±
0.5dgt.)/
℃
Voltage Measurement
Voltage Range
<0>6V
<1>60V
<2>100V
[UT3562]
<2>400V
[UT3563]
Maximum Display Value
±
6.06000V
±
60.6000V
±
101.000V
[UT3562]
±
404.000V
[UT3563]
Resolution
10
μ
V
100
μ
V
1mV
Precision
Slow
±
0.01%rdg.
±
3dgt
Medium
±
0.01%rdg.
±
5dgt
Fast
±
0.05%rdg.
±
5dgt
High- Speed
±
0.1%rdg.
±
6dgt
Temperature Index
(
±
0.001%rdg.
±
0.3dgt.)/
℃
Display: true and effective value, suitable for 10% to 100% of range.
Measurement conditions of the above data:
Temperature: 23
℃±
5
℃
Humidity:
≤
65%R.H.
Set Zero: Set Zero before Testing
Warm-Up Time: >60 Minutes
Calibration Time: Within 12 Months
Accuracy of Measurement Current: 10%