UT3500 series
User Manual
47
8.1.12 [Filter] Selection
The measurement stability of the instrument depends on the filter. Please select according
to the filter of the current region. If you are not sure, please select AUTO. The instrument
will automatically set the filter of the current region. The filter in China is 50 Hz.
Setting Steps:
1. Press the [Test] or [Setup] key, and select [SYSTEM CONFIG] with the function key at
the bottom of the screen to enter the <SYSTEM CONFIG> page;
2. Use the cursor keys to move the cursor to the [FILTER] field;
3. According to your own needs, press the function key at the bottom of the screen to
select AUTO or 50Hz or 60Hz.
8.1.13 [Default Setting]
After performing the default setting, all the settings of the instrument will be restored to the
parameters preset at the factory, including:
1. All the settings of the <SYSTEM CONFIG> page are restored to factory values:
2. The <SETUP> page is preset to factory values: