UT3500 series
User Manual
48
3. The <CATALOG> page is preset to file 0: the shutdown save is set to be prohibited:
8.2 System Info
Press the [Test] or [Setup] key, and select [SYSTEM CONFIG] with the function key at the
bottom of the screen to enter the <SYSTEM CONFIG> page; press the function key to
select [SYSTEM INFO]. This part contains the model, the serial NO. and FW version.
This page does not require the user to set up.
8.3 System Service
Press the [Test] or [Setup] key, and select [SYSTEM CONFIG] with the function key at the
bottom of the screen to enter the <SYSTEM CONFIG> page; press the function key to
select [SYSTEM SERVICE]. Warning: This page is not open to users. It is used to calibrate
data when leaving the factory. Non-professionals cannot enter by force. Otherwise, the
calibration data may be lost, resulting in a large deviation of the measurement data.
9. Handler Interface
This chapter mainly covers the following:
Terminals
Method of Connecting and Schematic Diagrams of Interfaces
The instrument provides users with a fully functional handler interface, which includes
qualified sorting output, HI/IN/LO, EOC (test completion signal), TRIG (external trigger
startup) input and other signals. Through this interface, the instrument can easily complete
the automatic control function with the user system control components.