UT3550 Battery Tester Users Manual
23
6.2.2
Example of Qualified Determination
Figure 6-2-2 Example of Qualified Determination Page
When the voltge or resistance measurement value is within the setting range of the comparator,
the judgment symbol
of the test value will be displayed on the screen, indicating that the
measured value is within the specification range. Only when the resistance and voltage are
within the range, the instrument will judge the total result of this group of test data as qualified,
and a qualified mark
will appear on the screen.
I
f the buzzer is set to “beep when qualified” at this time, the instrument will emit a qualified
buzzer for the qualified situation.
If the buzzer setting if OFF at this time, there will be no buzzer sound.
6.2.3
Example of Unqualified Determination
Figure 6-2-3 Example of Unqualified Determination
When the voltage or resistance measurement value exceeds or falls below the comparator
setting value, in any of these situations, the instrument determines that the test result is
unqualified, the symbol is
.
I
f the buzzer is set to “beep when unqualified” at this time, the instrument will emit a
unqualified buzzer for this situation.
If the buzzer setting if OFF at this time, there will be no buzzer sound.