background image

P/N:110401104397X

5. Rotary knob set to CONTINUITY: to measure low resistance.

6. Rotary knob set to 250V/500V/1000V:

    to select test voltage for insulation resistance measurement.

If low battery indicator shows on upper left corner of LCD after the 

meter is turned on, it means battery is almost used up and need to

be replaced.

Low battery indicator

(See Figure 1)

Figure 1

(1)  Set the rotary switch to ACV.

(2)  Insert the red test lead into "V" jack and the black test lead 

       into "G" jack.

Please take extreme caution in order to avoid electric shock when 

measuring high voltage.

Disconnect test leads and tested circuits and remove test leads 

away from input jacks after completing the measurement.

(See Figure 2)

lead

lead

test probe with the circuit  

under test.

With the instrument connected to tested circuit as described above,

Figure 2

(See Figure 3)

Figure 3

B

(See Figure 4)

Figure 4

figure 4,

then switch to any insulation measurement range,

Do not short-circuit the test leads under high-voltage output status 

or make insulation measurement after high voltage has already 

been output.

 

 Turn the knob to select test voltage 250V/500V/1000V.

lead

lead

With the knob already set to test voltage 250V/500V/1000V, then 

press TEST button,the instrument will be self-locked to measure 

continuously. The test voltage will be output and TEST button will 

light up. With the measurement finished, press TEST button to 

unlock and stop the measurement.

When red and black clips are connected to the circuit under test, 

high voltage is output from LINE jack and the current from 

EARTH jack.

Under insulation measurement mode, press PI/DAR button once 

to select PI parameter, the screen displays Time 1(1min)/Time2

(10mins), then press TEST to start PI measurement.

Cleaning the housing

Clean the instrument surface with soft cloth or sponge 

dampened with clean water.

To avoid damage to the instrument, do not submerge it into 

the water.

When it is necessary to verify or repair instrument, please

deliver the instrument to qualified professional serviceman

or designated repairing department.

the knob set to

batteries

installing the

Figure 5

wires

when replacing battery.

(See Figure 5)

Dielectric absorption ratio measurement

Under insulation measurement mode, press PI/DAR button two 

times to select DAR parameter, the screen displays Time 1(30s)/

Time2(1min), then press TEST to start DAR measurement, then 

press TEST three times to select another DAR, the screen shows 

Time 1(15s)/Time2(1min), press to begin another DAR 

measurement.

R

M

*END*

Reviews: