User’s Manual and UT5320R-SxA Series
23 / 43
Instruments.uni-trend.com
2)
Test result explaination of insulation resistance test
a) The instrument’s display format: if the display value is greater than 10 G, it will display as >10.0 GΩ.
b) The insulation value is read by the interface, the default unit is M. When then instrument displays >10.0GΩ,
it will read out the value which greater than 10 G. It’s only for reference.
5.6 Parameter Setting of Contact-Inspecting
(Only for UT5320R-S4A/UT5320R-S8A)
Figure 5-5
【
Meas Setup
】
3
Item
Input Range
Default Value Description
Voltage
(0.050~0.500) kV
0.100 kV
Output voltage during touch-inspecting test. Except in
special circumstances, use the default setting in
general.
LO-Limit
0.1uA
~
1.00 mA
500 uA
The lower limit of touch-inspecting alarm. Except in
special circumstances, use the default setting in
general.
Scan
Setup
ON,
OFF
OFF
UT5320R-S4A can control channel 1
,
2
,
3
,
4
;
UT5320R-S8A can control channel 1
,
2
,
3
,
4
,
5
,
6
,
7
,
8
;
5.7 Scanning Setting of Multi-Channel
(Only for UT5320R-S4A/UT5320R-S8A
)
Figure 5-6
【
Meas Setup
】
When UT5320R-S4A/S8A is in the test mode of AC withstand voltage, DC withstand voltage, insulation
resistance, which can configurate the scanning channel of multiple channels.