User’s Manual and UT5320R-SxA Series
29 / 43
Instruments.uni-trend.com
Address
0~32
1
If use Modbus (RTU) protocol, please set the station
address of this instrument:
①
The instrument allows use the station number
00 to brodcase communication.
②
1~32: The address for connecting the
instrument’s bus.
Result
FETCH?
,
auto
Auto
This function is only for SCPI protocol.
The instrument supports send data to the host
automatically. The data will be sent to the host
automatically after each test, and don’t need the
host to send FETCH? instruction.
InterLock
ON, OFF
OFF
The safety lock function prevents the test from
running unless the safety lock pin on the signal I/O
port connector is short-circuitted. Attached
pluggable terminal can be used for this purpose to
facilitate wiring. The detailes refer to Chapter 8.2.
IP addr
192.168.001.123
When communication mode is TCP, use numeric
keyboard to change the parameter.
Port
502
It is the default value and cannot be change. It will
be used when the communication mode is TCP.
6.2.1 Reset to factory setting
All settings of the instrument will reset to the factory setting.
<File Managment> page is preset to file 1.
6.3 <System Information>
Enter <SYSTEM 2> page, press function key to select [System Information].
The information includes model name, serial number and version of the instrument.
This page don’t need to setup.
Figure 6-3 <System Information> Page