User’s Manual and UT5320R-SxA Series
39 / 43
Instruments.uni-trend.com
Output
Accuracy of
Voltage
±(1.0% S5V) (no load)
±(1.0% Setting +5V)
(no load)
Test
Accuracy of
Voltage
±(1.0% R5V)
±(1.0% R5V)
Producing
Mode
of
Voltage
DDS signal source plus AB power
amplifier
DDS signal source
plus
AB
power
amplifier
Current
Test Range
AC
Current
Range
0.001mA – 10mA
0.001mA – 20mA
Short-
circuit
Current
>20 mA
>40 mA
(Transient
)
(output voltage setting >500V)
(output
voltage
setting >500V)
Resoultion
of Current
0.001 mA
0.001 mA
Accuracy of
Current
±(2.0% R 5 characters)
±(2.0% Reading + 5
characters)
Actucal
Current
OFF, 0.001 mA-10mA
OFF, 0.001 mA-20mA
DC
Current
Range
0.1uA – 5.00mA
0.1uA – 10.00mA
Accuracy of
Current
±(2.0% R 5 characters)
±(2.0% Reading + 5
characters)
Discharge
Automaically discharge after the
test is finished
Automaically
discharge after the
test is finished
Insulation Resistance Test
Output Voltage
0.050kV–2.500kV
(optional 5.00kV
)
0.050kV–
2.500kV(optional
5.00kV
)
Rsoulation of Voltage
1V
1V
Test Accuracy of Voltage
±(1.0% R2V)
±(1.0% R2V)
Maximum Output Current
5mA
10mA
Maximum Output Power
12.5VA (2500V/5mA)
25VA (2500V/10mA)
Output Transient Short-circuit
Current
>10mA (output voltage
setting >500V)
>20mA(output voltage
setting >500V)