1. Safety Regulation
This chapter includes the following contents.
Workstation Arrangment
Operator Regulation
Prohibitied Operation
Dayily Inspection for guarantee long-term use without failures
1.1 Workstation Arrangment
Position of Workstation
Workstation should arrange in a capacious filed that is not necessary for general personnel, so that non-staff
members are far away from the workstation. The workstation must separate from other facilities and it need to
indicated particularly “High Voltage Test Worksation”. During the testing, it must indicate “DANGER! Stay Away!”
Input Power
Be sure to connect the instrument to ground for keep safe. The power supply of the workstation should have a
separate siwth. The switch need to install in the prominent position on the entrance and marked with special
sign to let people know this is the power switch of the workstation. Once an emergency accident happens, the
switch can be shut down immediately and then deal with the accident.
Workplace
An insulated worktable or workbench is must, any mental can not be used between operator and DUT. When
design the workdfiled, it is not allowed that the operator across the DUT to operating the test instrument. The
workplace should be tidy and clean. Unused instrument and test wire should put in a fixed position, make sure
that all personnel can immediately distinguish the parts under test, to be tested and tested. The test station and
its surrounding can not have flammable and corrosive gas and the tester can not be used near the flammable
substance.
1.2 Operator Regulation
Qualification of Personnel
The operating of this series tester is dangerous, misoperation will cause injury to personnel, and this injury is
even life-threatening. So the user must be trained and strictly follow the user’s manual.
Safety Regulation
The operator must be given safety education and training at any time to understand the importance of various
safety operations and follow the safety regulation to operating the tester.
Dress Code
The operator shall not wear clothes with metal decoration, metal bracelet and watch, these metal ornaments
may cause accidental electrical shock. The operator must wear the insulation gloves when operating the tester.
Medical Regulation
It must not be allowed the personnel who has heart disease or wear heart rate regulator and cardiac pacemaker
to operating the tester.