22
Note:
l
The test leads can add 0.1
Ω
to 0.2
Ω
of error to
resistance measurement. To obtain precision readings
in low-resistance measurement (
UT60F
: 400
Ω
and
UT60G
is 600
Ω
), short-circuit the input terminals
beforehand and record the reading obtained (called
this reading as X).
Then use the equation:
measured resistance value (Y) –(X) = precision
readings of resistance.
l
For high-resistance measurement (>1M
Ω
), it is normal
taking several seconds to obtain a stable reading.
l
If
Ω
reading with shorted test leads is not 0.5
Ω
,
check for loose test leads, wrong function selected,
or enabled data hold function.
l
The LCD displays
OL
indicating open-circuit or the
tested resistor value is higher than the maximum
range of the Meter.
l
Resistance measurement is default to auto range
mode. At
Ω
range, open circuit voltage approximate
1.20V. At all other range, open circuit voltage
approximate 0.45V.
l
When resistance measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test and remove testing leads
from the input terminals.
Measurement Operation(6)
Model UT60F/G: OPERATING MANUAL