33
Accuracy Specifications(2)
C. DC Current
0.1
µ
A
1
µ
A
10
µ
A
100
µ
A
10mA
Remarks
l
Overload protection:
Model UT60F at 400.0
µ
A to 400.0mA range:
0.5A, 250V, fast type fuse, ø5×20mm
Model UT60F at 10.00A range
:
10A, 250V, fast type fuse, ø5×20mm
Model UT60G at 60.00mA to 600.0mA range:
1A, 250V, fast type fuse, ø5×20mm
Model UT60G at 10.00A range:
10A, 250V, fast type fuse, ø5×20mm
l
At 10.00A range:
For continuous measurement 10 seconds and interval
time between 2 measurements greater than 15 minutes.
Range
Resolution
Accuracy
UT60F UT60G
400.0
µ
A
4000
µ
A
40.00mA
400.0mA
10.00A
60.00mA
600.0mA
10.00A
(1%+2)
(1.2%+3)
(1.5%+5)
D. AC Current
Remarks
l
Overload protection:
Model UT60F at 400.0
µ
A to 400.0mA range
:
0.5A, 250V, fast type fuse, ø5×20mm
Range
Resolution
Accuracy
400.0
µ
A
4.000mA
40.00mA
400.0mA
10.00A
60.00mA
600.0mA
10.00A
(1.5%+5)
(2%+5)
(2.5%+5)
UT60F UT60G
0.1
µ
A
1
µ
A
10
µ
A
100
µ
A
10mA
Model UT60F/G: OPERATING MANUAL