19
¡
D. Measuring Resistance / Conductance,
Capacitance & Continuity
To avoid damages to the Meter or to the devices
under test, disconnect circuit power and discharge
all the high-voltage capacitors before measuring
resistance, capacitance & continuity. Use the DC
voltage function to confirm that the capacitor is
discharged.
Resistance / Conductance Measurement
(see figure 6)
The resistance ranges are: 800.00
¡
, 8.0000k
¡
,
80.000k
¡
, 800.00k
¡
, 8.0000M
¡
, 80.000M
¡
. To
measure the resistance, Conductance, connect the
Meter as follows:
1. Insert the red test lead into the terminal
and the black test lead into the
COM
terminal.
2. Set the rotary switch to ; resistance
measurement is default.
3. Connect the test leads across with the object
being measured. The measured value shows on
the display.
¡
¡
¡
¡
¡
¡