Before measuring current, check the Meter's fuses
and turn off power to the circuit before connecting
the Meter to the circuit.
Replace the battery as soon as the battery indicator
appears. With a low battery, the Meter might
produce false readings that can lead to electric
shock and personal injury.
Remove test leads and RS232C interface cable
from the Meter and turn the Meter power off before
opening the Meter case.
When servicing the Meter, use only the same model
number or identical electrical specifications
replacement parts.
The internal circuit of the Meter shall not be altered
at will to avoid damage of the Meter and any accident.
Soft cloth and mild detergent should be used to
clean the surface of the Meter when servicing. No
abrasive and solvent should be used to prevent the
surface of the Meter from corrosion, damage and
accident.
The Meter is suitable for indoor use.
Turn the Meter off when it is not in use and take
out the battery when not using for a long time.
Constantly check the battery as it may leak when
it has been using for some time, replace the battery
as soon as leaking appears. A leaking battery will
damage the Meter.
Rules For Safe Operation(2)
Summary of Contents for UT70D
Page 8: ...The Meter Structure see figure 1 ...
Page 12: ...Display Symbols 1 see figure 2 No Symbol Meaning 8 6 4 3 ...
Page 13: ......
Page 34: ...34 Turning on the Display Backlight ...
Page 35: ...35 ...
Page 41: ...41 5 Duty Cycle 1 99 600Vp ...
Page 43: ...43 MAINTENANCE 2 ...
Page 44: ...44 MAINTENANCE 3 ...
Page 46: ...46 B Setting of RS232C Serial Ports see figure 15 ...