Figure 6
Replace the fuses using the following procedure:
1. Remove the test leads from the calibrator
terminals and turn the calibrator off.
2. Remove the screw on the battery door, then
remove the battery door and take out the
battery.
3. Remove the three screws from the case bottom
and turn the case over.
4. Gently remove the fuse from its mounting
bracket.
5. Replace the blown fuse with a 125mA 250V
fuse.
6. Fit the top and bottom covers together.
7. Reinstall the three screws.
8. Replace the battery door and reinstall the
battery door screw.
Model UT713: OPERATING MANUAL