64
Model UT71A/B/C/D/E: OPERATING MANUAL
Range
Resolution
400µA
4000µA
40mA
400mA
10A
Bandwidth
Accuracy
Overload
Protection
UT71C/D/E
UT71C/D/E
UT71A/B
200µA
2000µA
20mA
200mA
10A
UT71A
UT71B
D. AC Current (AC+DC measurement is available)
0.01µA
0.1µA
0.001mA
0.01mA
0.001A
45Hz~1kHz
1kHz~10kHz
45Hz~1kHz
1kHz~10kHz
(0.7%+15)
(1%+40)
(1.5%+20)
(5%+40)
0.5A, 1000V,
fast type fuse,
6.3x32mm
10A,1000V, fast type
fuse, 10.3x38mm
Remarks:
l
l
Display:
a)True rms are valid from 10% of range to 100% of range
b) AC crest factor can be up to 3.0.
c) A residual reading of 80 digits with test leads shorted, will not affect stated accuracy.
d) When frequency is lower than 100kHz, the accuracy guarantee range 10%-100%.
e) When making AC+DC measurment, the accuray need to add (1%+ 35 digits) of reading based on the above table.
At 10A range:
a) When the measured current is 5A, continuous measurement is allowed.
b) When the measured current is between 5A-10A, continuous measurement 10 seconds and interval more
than 15 minutes.
(1%+15)
(2%+40)
(2%+20)
(6%+40)
(0.8%+15)
(1.5%+40)
(2%+20)
(6%+40)