Model UT805: OPERATING MANUAL
B. AC Voltage
(750V range: frequency 1kHz, all other
ranges: frequency 20kHz)
200mV:
2V :
20V :
200V :
Input 19mV, press
CAL
button to display
(-- CAL--) (-LO-END)
(
) Input
190mV, press
CAL
button to display (--CAL--)
( - H I - E N D )
( 1 9 0 . 0 0 0 m V ) ,
calibration finished
Input 190mV, press
CAL
button to display
(--CAL--)
(-LO-END)
(
) Input
1.9V, press
CAL
button to display (-- CAL--)
(-HI-END)
(1.90000V), calibration
finished.
Input 1.9V, press
CAL
button to display
(-- CAL--) (-LO-END)
(
) Input
19V, press
CAL
button to display (-- CAL--)
(-HI-END)
(19.0000V), calibration
finished.
Input 19V, press
CAL
button to display
(--CAL--)
(-LO-END)
(
) Input
190V, press
CAL
button to display (-- CAL--)
(-HI-END)
(190.000V), calibration
finished.
750V :
AC+DC:
Input 190V, press
CAL
button to display
(-- CAL--) (-LO-END)
(
) Input
750V, press
CAL
button to display (-- CAL--)
(-HI-END)
(750.00V), calibration
finished.
Carry out the calibration according to each
range of DC voltage.