UT8805E Benchtop Digital Multimeter
26
Turn the relative operation on, the displayed value is actual measured value minus the set relative value.
About relative value setting, please refer to 2.10 Mathematical Functions.
5)
Unit setting
3 units of temperature, °C, °F, K, are selectable in UT8805E.
6)
Test the testing point and read the displayed value.
Note:
When the input terminal is in dangling state in thermocouple temperature measurement, random
reading will show.
2.3.11 Capacitance Measurement
The maximum capacitance range is 2mF. The connecting and measurement methods are shown below:
Steps:
1)
Press
button to enter capacitance measuring interface as Figure 2-17.
Figure 2-17 capacitance measuring interface
2)
Connect the test leads to the measured capacitance as Figure 2-18, the red test lead to the terminal
Input-HI and the capacitance anode, and the black one to terminal Input-LO and cathode.
Figure 2-16 Connection of 4-wire resistance measurement
3)
Select proper range according to the measured capacitance.
Range
2nF, 20nF, 200nF, 2μF, 20μF, 200μF, 2mF
Input protection
DC 1000V or AC 750Vrms for all ranges (terminal
HI)
Configurable parameters
Range, relative operation set value
Note:
all ranges are allowed to exceed range by 20%.
4)
Relative value setting
Turn the relative operation on, the displayed value is actual measured value minus the set relative value.
About relative value setting, please refer to 2.10 Mathematical Functions.
5)
Measured value reading