UT8805E Benchtop Digital Multimeter
14
2.2 Reading Speed Selection
There are 3 reading speeds can be selected by soft key menu: [Slow] 2.5 reading/s, [Medium] 10
reading/s and [Fast] 5k reading/s. The display refreshing rate is 10 times/s. Press [Speed] button, and
then select [Slow], [Medium] or [Fast] to select measuring speed.
Figure 2-3 Interface of measuring speed selecting
Note:
1) There are 3 reading speeds (slow, medium and fast) can be set in DCV, DCI and 2W/4W resistance
functions.
2) In slow speed measurement, the zero reading function of DCV, DCI and 2W/4W resistance is on,
which can restrain thermo-electromotive force and offset voltage/current caused by temperature
variation and component aging.
3) 2.5 reading/s and 10 reading/s are corresponding to 5.5 bits reading resolution.
4) 5k reading/s is corresponding to 5.5 bits reading resolution.
5) Temperature measurement is corresponding to 5.5 bits reading resolution (slow speed).
6) Diode/continuity measurement is corresponding to 4.5 bits reading resolution (medium speed).
7) Frequency measurement is corresponding to 5.5 bits reading resolution, and the reading speed
depends on the input signal frequency.
8) Capacitance measurement is corresponding to 3.5 bits reading resolution, and the reading speed
depends on the charge-discharge time of capacitor.
2.3 Basic Measurements
The basic measurement functions of UT8805E include measurements of DC/AC voltage, DC/AC current,
2/4-wire resistance, capacitance, continuity, diode, frequency/cycle and temperature.
2.3.1 DC Voltage Measurement
The maximum DC voltage range of UT8805E is 1000V, and DC voltage measurement is selected by
default. The connecting methods and operation are introduced below.
Steps:
1) Press
button to enter interface of DC voltage measurement as Figure 2-4.