UTG1000A Series                                                                             29 

 

 

4.1.2 Frequency Modulation (FM) 

In frequency modulation, modulated waveform is usually composed of carrier wave and modulation shape. Carrier 
wave frequency will change as the amplitude of modulation shape changes. 

Press Menu

Modulation

Type

→ 

Frequency Modulation in turn to start the FM function. The device will output 

modulated waveform with modulation waveform and carrier wave set currently. 

 

 

Carrier Wave Waveform Selection 

FM carrier waveform can be: Sine wave, square wave, ramp wave, pulse wave, arbitrary wave (except DC) and noise 
(the default is sine wave). After selecting FM modulation, press Carrier Wave Parameter softkey to enter carrier 
waveform selection interface. 

 

Summary of Contents for UTG1000 Series

Page 1: ...User Manual UTG1000 Series Function Arbitrary Waveform Generator 2016 09 UNI T Technologies Inc...

Page 2: ......

Page 3: ...ni Trend device In order to use this instrument safely and correctly please read this manual thoroughly especially the Safety Notes part After reading this manual it is recommended to keep the manual...

Page 4: ...anty service The customer shall be responsible for packing and shipping the defective products to the designated maintenance center of UNI T pay the shipping cost and provide a copy of the purchase re...

Page 5: ...r UNI T shall pay the return shipping fee If the product is sent to any other location the customer shall be responsible for all shipping duties taxes and any other expenses This warranty shall not ap...

Page 6: ...sure that the product is properly grounded before connecting to the input or output of the product Correct connection of oscilloscope probe Ensure that the probe ground and ground potential are correc...

Page 7: ...may cause damage to the product and other properties The following terms may appear on the product Danger Performing this operation may cause immediate damage to the operator Warning This operation ma...

Page 8: ...es 15 3 1 3 Machine Inspection 15 3 2 Handle Adjustment 15 3 3 Basic Waveform Output 16 3 3 1 Frequency Setting 16 3 3 2 Amplitude Setting 16 3 3 3 DC Offset Voltage Setting 17 3 3 4 Square Wave Setti...

Page 9: ...0 4 3 Arbitrary Wave Output 74 4 3 1 Enable Arbitrary Wave Function 74 4 3 2 Arbitrary Wave Selection 74 Chapter 5 Trouble Shooting 76 5 1 No Display On Screen Black Screen 76 5 2 No Waveform Output 7...

Page 10: ...diate damage to the operator Warning This operation may cause potential damage to the operator Note This operation may cause damage to the product and devices connected to the product Symbols on the p...

Page 11: ...ing to the input or output of the product To avoid personal injury and prevent damaging the product only trained personnel can perform the maintenance program To avoid fire or electric shock please no...

Page 12: ...Square wave pulse waveform of 5MHz and its rising falling and duty cycle time are adjustable Using DDS implementation method with 125M s sampling rate and 14bits vertical resolution 6 bit high precisi...

Page 13: ...Button This button includes 3 function labels Channel settings frequency meter and system A highlighted label the midpoint of the label is gray and font is pure white has a corresponding sub label at...

Page 14: ...multifunctional knob as direction button Press the multifunctional knob to select function set parameters and confirm selection 11 Number Keyboard Number keyboard is used to enter parameter number 0 t...

Page 15: ...erface is shown in figure 2 3 Figure 2 3 Detailed Description Channel information 1 ON OFF on the left is channel open information 2 There is a Limit logo indicates output range limit where white is v...

Page 16: ...reen Sub label contents belongs to the next category of Type label Press corresponding button to select sub labels Waveform Parameter List Displays parameters of current waveform in a list Waveform Di...

Page 17: ...distributor will arrange for repairment or replacement 3 1 2 Check Accessories UTG1000 accessories are Power cord USB data cable BNC cable 1 meter and user CD If any of the accessories are missing or...

Page 18: ...setting mode Set parameters by pressing Frequencysoftkey to change frequency and period b Use number keyboard to input the required number of 2 5 c Select corresponding unit MHz 3 3 2 Amplitude Settin...

Page 19: ...e default waveform is a sine wave with 0V DC offset voltage with 50 termination Steps for changing DC offset voltage to 150mV are shown as following 1 Press Menu Waveform Parameter Offsetto enter para...

Page 20: ...h 2ms period 1 5Vpp amplitude 0V DC offset and 25 duty cycle limited by the minimum pulse width specification 80ns 200us rising time and 200us falling time are seen as following Press Menu Waveform Ty...

Page 21: ...be set by multifunctional knob and direction buttons 3 3 7 Ramp Wave Setting Default symmetry degree of ramp wave is 100 Steps for setting triangular wave with 10kHz frequency 2V amplitude 0V DC offse...

Page 22: ...it Note This parameter can be set by multifunctional knob and direction buttons 3 4 Frequency Measurement This device is suitable for measuring frequency and duty cycle of TTL compatible signals with...

Page 23: ...are shown as following Long press any softkey or button to display relevant information If the content is more than 1 screen size use softkey or multifunctional knob to display the next screen Press...

Page 24: ...to start the AM function Then the modulated waveform will output with modulation waveform and carrier wave set Carrier Waveform Selection AM carrier waveform can be sine wave square wave ramp wave or...

Page 25: ...value and select unit after selecting carrier waveform Modulation Source Selection This device can select internal modulation source or external modulation source After enabling AM function the defau...

Page 26: ...ting range of AM modulation depth is 0 to 120 and the default is 100 When modulation depth is set to 0 the constant amplitude a half of the carrier wave amplitude that has been set is output Output am...

Page 27: ...5 2 Set Modulation Signal Parameter After enabling the AM function press Parametersoftkey and the interface will appear as following Press corresponding softkey then enter required numerical value and...

Page 28: ...Press Carrier Wave Parameter Type Square Wave in turn to select square wave as carrier wave signal Press Parametersoftkey again and the interface will pop up as following Press corresponding softkey t...

Page 29: ...etting carrier wave parameter press Returnsoftkey to back to the following interface for setting modulation depth Press Parameter Modulation Degreesoftkey again then enter number 80 and press softkey...

Page 30: ...nel Setupsoftkey after pressing Utility button and popping up labels After channel output is opened backlight of Channel button is on and on the right side of channel information label the font OFF ch...

Page 31: ...odulation Type Frequency Modulation in turn to start the FM function The device will output modulated waveform with modulation waveform and carrier wave set currently Carrier Wave Waveform Selection F...

Page 32: ...alue Maximum Value Minimum Value Maximum Value Sine Wave 1 Hz 10MHz 1 Hz 10MHz 1 Hz 5MHz Square wave 1 Hz 5MHz 1 Hz 5MHz 1 Hz 5MHz Ramp Wave 1 Hz 400kHz 1 Hz 400kHz 1 Hz 400KHz Arbitrary Wave 1 Hz 3MH...

Page 33: ...n negative signal level FM output frequency is less than carrier wave frequency Low external signal level has small deviation For example if the frequency offset is set to 1kHz and the external modula...

Page 34: ...ave frequency is less than the allowed maximal frequency of current carrier wave If the frequency deviation value is set to an invalid value the device will automatically set the offset value to the c...

Page 35: ...3 Press corresponding softkey then enter required numerical value and select the unit 3 Set Carrier Wave Signal Parameter Press Carrier Wave Parameter Type Sine Wave in turn to select sine wave as car...

Page 36: ...34 UTG1000A Series Press Parametersoftkey and the interface will pop up as following Press corresponding softkey first then enter required numerical value and select the unit 4 Set Frequency Deviation...

Page 37: ...ter press Returnsoftkey to back to the following interface for setting frequency deviation Press Parameter Frequency Deviation softkey then enter number 5 and press kHzsoftkey with number keyboard for...

Page 38: ...dulation PM In phase modulation modulated waveform is usually composed of carrier wave and modulation wave The phase of carrier wave will change as the amplitude of modulation shape changes Press Menu...

Page 39: ...e carrier wave frequency range is different of different carrier waveform Default frequency of all carrier wave is 1kHz The frequency setting range of each carrier wave can be seen in the following ta...

Page 40: ...er Wave Parameter Type in turn 2 External Source When modulation source is external carrier waveform will be modulated by an external waveform PM phase deviation is controlled by 5V signal level of ex...

Page 41: ...then set a sine wave with 200Hz from the internal of the instrument as a modulation signal and a square with frequency of 900Hz and amplitude of 100mVpp as a carrier wave signal Finally set the phase...

Page 42: ...00A Series 3 Set Carrier Wave Signal Parameter Press Carrier Wave Parameter Type Sine Wave in turn to select sine wave as carrier wave signal Press Parametersoftkey and the interface will pop up as fo...

Page 43: ...000A Series 41 Press corresponding softkey then enter required numerical value and select the unit 4 Set Phase Deviation Press Returnsoftkey to back to the following interface for setting phase modula...

Page 44: ...ey then enter number 200 and press softkey with number keyboard for setting phase deviation 5 Enable Channel Output Press Channel button to open channel output quickly The shape of PM modulation wavef...

Page 45: ...ation signal ASK Modulation Selection Press Menu Modulation Type Amplitude Shift Keying in turn to start the ASK function the device will output modulated waveform with ASK rate and carrier wave set c...

Page 46: ...ue Maximum Value Minimum Value Maximum Value Sine Wave 1 Hz 10MHz 1 Hz 10MHz 1 Hz 5MHz Square wave 1 Hz 5MHz 1 Hz 5MHz 1 Hz 5MHz Ramp Wave 1 Hz 400kHz 1 Hz 400kHz 1 Hz 400KHz Arbitrary Wave 1 Hz 3MHz...

Page 47: ...an the amplitude of current setting when external input logic is high ASK Rate Setting When modulation source is internal frequency of ASK amplitude jump can be modulated After enabling ASK function A...

Page 48: ...46 UTG1000A Series 2 Set Carrier Wave Signal Parameter Press Carrier Wave Parameter Type Sine Wave in turn Press Parametersoftkey and the interface will pop up as following...

Page 49: ...ue and select the unit 3 Set ASK Rate After setting carrier wave parameter press Returnsoftkey to go back to the following interface for setting phase modulation Press Parameter Ratesoftkey again then...

Page 50: ...48 UTG1000A Series 4 Enable Channel Output Press Channel button to open channel output quickly The shape of ASK modulation waveform checked through oscilloscope is shown as following...

Page 51: ...with current setting Carrier Wave Waveform Selection Press Carrier Wave Parametersoftkey to enter carrier waveform selection interface FSK carrier waveform can be sine wave square wave ramp wave or a...

Page 52: ...abling FSK function the default of modulation source is internal If need to change press Parameter ModulationSource External in turn 1 Internal Source When modulation source is internal internal modul...

Page 53: ...frequency between carrier wave frequency and hop frequency can be set After enabling FSK function FSK rate can be set and the settable range is 2mHz to 100kHz the default rate is 1kHz If need to chang...

Page 54: ...000A Series Press Parametersoftkey again and the interface will pop up as following Press corresponding softkey first then enter required numerical value and select the unit 3 Set Hop Frequency and FS...

Page 55: ...Press Returnsoftkey to go back to the following interface Press Parametersoftkey again and the interface will pop up as following Press corresponding softkey first then enter required numerical value...

Page 56: ...n generator can be configured to move between two preset phase carrier wave phase and modulation phase Output carrier wave signal phase or hop signal phase on the basis of the logic of modulation sign...

Page 57: ...Settable carrier wave frequency range is different of different carrier waveform Default frequency of all carrier wave is 1kHz The frequency setting range of each carrier wave can be seen in the foll...

Page 58: ...lation phase 2 External Source When modulation source is external carrier waveform will be modulated by an external waveform Carrier wave phase will be output when external input logic is low and modu...

Page 59: ...dulation phase move between each other with 1kHz frequency Specific steps are seen as following 1 Enable Phase Shift Keying PSK Function Press Menu Modulation Type Phase Shift Keying in turn to start...

Page 60: ...58 UTG1000A Series Press corresponding softkey then enter required numerical value and select the unit 3 Set PSK Rate and Modulation Phase Press Returnsoftkey to go back to the following interface...

Page 61: ...ss Parametersoftkey and the interface will pop up as following Press corresponding softkey then enter required numerical value and select the unit 4 Enable Channel Output Press Channel button to open...

Page 62: ...lated waveform is usually composed of carrier wave and modulation shape and the pulse width of carrier wave will change as modulation shape amplitude changes PWM Modulation Selection Press Menu Modula...

Page 63: ...0uH to 25MHz and the default frequency is 1kHz Press Parameter Frequency softkey to change frequency then enter required numerical value and select unit Carrier Wave Duty Cycle Setting Settable range...

Page 64: ...the default of modulation wave frequency is 1kHz If need to change press Carrier Wave Parameter Modulation Frequency in turn When modulation source is external carrier wave waveform pulse wave will b...

Page 65: ...d a pulse wave with 10kHz frequency 2Vpp amplitude and 50 duty cycle as a carrier wave signal finally set duty cycle deviation to 40 Specific steps are seen as following 1 Enable Pulse Width Modulatio...

Page 66: ...TG1000A Series 3 Set Carrier Wave Signal Parameter Press Carrier Wave Parameter softkey to enter carrier wave parameter setting interface Press Parameter softkey and the interface will pop up as follo...

Page 67: ...required numerical value and select the unit 4 Set Duty Cycle Deviation Press Returnsoftkey to back to the following interface for duty cycle deviation setting After pressing Parameter Dutycyclesoftke...

Page 68: ...66 UTG1000A Series 5 Enable Channel Output Press Channel button to open channel output quickly The shape of PWM modulation waveform checked through oscilloscope is shown as following...

Page 69: ...d sine wave square wave ramp wave and arbitrary wave except DC can produce sweep output 4 2 1 Sweep Selection 1 Enable Sweep Function Press Menu button first then press Sweepsoftkey to start sweep fun...

Page 70: ...equivalent to stop frequency DDS function generator sweeps output fixed frequency Synchronous signal of sweep mode is a signal that is low from the start of sweep time to the middle of sweep time and...

Page 71: ...he default is 1s and the settable range is from 1ms to 500s If need to change press Parameter Sweep Timesoftkey in turn then enter number with number keyboard and press corresponding unit softkey 4 2...

Page 72: ...l modulation interface Input CNT probe at the beginning of sweep In manual trigger signal generator output a pulse that has pulse width more than 1us through external modulation interface Input CNT pr...

Page 73: ...71 2 Select Sweep Waveform Press Carrier Wave Paremeter Type Square Wavesoftkey to select sweep waveform and the interface will pop up as following Press Parametersoftkey and the interface will pop u...

Page 74: ...then enter required numerical value and select the unit 3 Set Initial Terminal Frequency Sweep Time Trigger Source and Trigger Edge Press Returnsoftkey to the following interface Press Parametersoftk...

Page 75: ...UTG1000A Series 73 Press corresponding softkey then enter required numerical value and select the unit 4 Enable Channel Output Press Channel button to open channel output quickly...

Page 76: ...ound in table 4 1 built in arbitrary wave list 4 3 1 Enable Arbitrary Wave Function Press Menu Waveform Type Arbitrary Wave in turn to start the arbitrary wave function The device will output arbitrar...

Page 77: ...5 Table 4 1 Built in Arbitrary Wave List AbsSine AmpALT AttALT Gaussian Monopulse GaussPulse SineVer StairUd Trapezia LogNormalSinc Sinc Electrocardiogram Electroencephalogram Index Rises Index Falls...

Page 78: ...ethod press Utility System System About in turn 5 1 No Display On Screen Black Screen When the power button is pressed and the oscilloscope is black screen a Check the power supply connection b Make s...

Page 79: ...any other explicit or implied guarantee including but not limited to the product trading and special purpose for any implied warranties In any case UNI T for indirect special or consequential loss do...

Page 80: ...Amplitude Lower Limit 5V Basic Wave Frequency 1kHz Ampltide 100mVpp DC Offset 0mV Initial Phase 0 Duty Cyle of Square Wave 50 Symmetryof Ramp Wave 100 Duty Cycle of Pulse Wave 50 Lead Edge of Pulse W...

Page 81: ...rnal Modulation Shape Pulse Wave Modulation Frequency 100Hz Duty Cycle Deviation 20 ASK Modulation Modulation Source Internal ASKRate 100Hz FSK Modulation Modulation Source Internal Carrier Wave Frequ...

Page 82: ...80 UTG1000A Series Parameters of System Sound of Buzzer Open Number Format Backlight 100 Language Determined by Factory Settings...

Page 83: ...M PM ASK FSK PSK PWM Features of Waveform Sine Wave Frequency Range 1 Hz 20M Hz 1 Hz 10M Hz 1 Hz 5MHz Resolution 1 Hz Accuracy 50ppm in 90 days 100ppm in one year 18 C 28 C Harmonic Distortion Typical...

Page 84: ...Range Peak Value AC DC 5V 50 10V High Resistance Offset Precision 1 of offset setting 0 5 of ampltide 2mV Features of Arbitrary Waveform Frequency Range 1 Hz 3MHz 1 Hz 2MHz 1 Hz 1MHz Resolution 1 Hz W...

Page 85: ...ion Depth 0 120 FM Modulation Carrier Wave Sine Wave Square Wave Ramp Wave Arbitrary Wave Source Internal External Modsulation Shape Sine Wave Square Wave Ramp Wave Noise Arbitrary Wave Modulation Fre...

Page 86: ...Pulse Wave Source Internal External Modulation Shape Sine Wave Square Wave Ramp Wave Noise Arbitrary Wave Modulation Frequency 2mHz 50kHz Width Deviation 0 49 99 of pulse width Sweep Carrier Wave Sine...

Page 87: ...l display Display Resolution 480 horizontal 272 vertical Power Power Voltage 100 240 VAC 45 440Hz CAT II Consume Power Less than 25W Fuse 2A T Level 250V Environment Temperature Range Operation 10 40...

Page 88: ...86 UTG1000A Series Appendix C Accessories List Type UTG1000A Standard Accessories Power line meets local country standard USB data cable UT D06 BNC cable 1 meter User CD Warranty card...

Page 89: ...damp but not dripping soft cloth wipe the instrument it is suitable to use mild cleaning agent or water to wipe the dust on instrument do not use chemistry or cleaning agent with potent substances lik...

Page 90: ...88 UTG1000A Series This user manual may be revised without prior notice...