4
T
08 9351 7500
E
info@unicarehealth.com.au
W
www.unicarehealth.com.au
Jib
Spreader Bar
Actuator
Push Handle
Mast
Base Chassis
Leg
Battery &
Control Box
Page 1: ...KRATOS 320 Mobile Patient Lifting Hoist Instruction Safety Manual Please read this manual before operation for maintenance instructions and safe usage The enabling power of applied knowledge...
Page 2: ...ts Overview 2 Important Saftey Considerations 3 Assembly Instructions 5 Hoist Operation 6 Actuator 7 Charging 7 Charging Indicator Display 7 External Charging Unit 8 Charging Do s and Don ts 8 Lifting...
Page 3: ...rs The swivel feature of the detachable yoke allows a patient to be rotated through 360 at full lift height thus making the carers job easier and safer The KRATOS 320 lifting range allows patients to...
Page 4: ...ng The lifting transferring task should be assessed planned by a trained user SAFETY CONSIDERATIONS Ensure that all users read this guide before operating the hoist Do not operate the hoist without tr...
Page 5: ...4 T 08 9351 7500 E info unicarehealth com au W www unicarehealth com au Jib Spreader Bar Actuator Push Handle Mast Base Chassis Leg Battery Control Box...
Page 6: ...nt between mast and base chassis 3 Attach lift actuator to jib by aligning actuator clevis with actuator mount bracket Insert bolt and tighten nylock nut 4 Attach push handle assembly with bolts suppl...
Page 7: ...released the jib will remain in position The KRATOS 320 hoist incorporates an actuator driven leg opening adjustment Two buttons will be seen on the handset with arrows and symbolised legs indicating...
Page 8: ...harger This is powered by plugging the 3 pin plug into a standard 240V ac outlet To recharge the battery pack perform the following steps 1 Release battery pack from control box by pulling the handle...
Page 9: ...stow the handset and if tted with charger the mains power cable when transporting the hoist Do clean the actuators control box charger battery and handset at regular intervals to remove dust and dirt...
Page 10: ...aining manual but as a general guide to lifting procedure only Fit sling to patient REFER TO SLING INSTRUCTION FOR USE Manoeuvre the hoist around the patient adjusting the legs to t around chair wheel...
Page 11: ...castor brakes operate ef ciently Over haul castors pack with grease Inspect surfaces for scratches chips Clean surfacers with a neutral detergent remove any adhesive tape with orange or eucalyptus oil...
Page 12: ...t up No mains connection Internal fuse blown Battery chargers is defective Plug into mains Replace fuse Repair replace charger Actuator will not run and no relay click is heard from control box Contro...
Page 13: ...lectric Adjustment Base height under bed clearance 155mm Minimum height of spreader bar 575mm Maximum height of spreader bar 2065mm Hoisting range 1490mm Lifter weight 61kg Patient lift space 1500 Und...
Page 14: ...the improper or unsafe use of the product No warranty claim shall apply where the product or any part thereof has been modi ed varied altered or damaged either accidentally or through improper or neg...
Page 15: ...ing slings in both mesh and fabric to suit the ALLEGRO range of patient lifting hoists A range of custom made to speci cation slings are available on request Sling Sizes Available X small small medium...
Page 16: ...OT SLINGS SWL Safe working load this symbol represents the maximum load the lifter sling or attachment is rated to for safe operation SLING CARE SYMBOLS SLING MAINTENANCE AND CARE WARNING Prior to use...
Page 17: ...7500 E info unicarehealth com au W www unicarehealth com au Log Book Service Type Condition Report Date Inspection by Signature Action Taken Service Type Condition Report Date Inspection by Signature...
Page 18: ...truction Safety Manual T 08 9351 7500 E info unicarehealth com au W www unicarehealth com au Log Book Service Type Condition Report Date Inspection by Signature Action Taken Service Type Condition Rep...
Page 19: ...7500 E info unicarehealth com au W www unicarehealth com au Log Book Service Type Condition Report Date Inspection by Signature Action Taken Service Type Condition Report Date Inspection by Signature...
Page 20: ...PO Box 82 Bentley WA 6982 T 08 9351 7500 F 08 9350 5299 info unicarehealth com au www unicarehealth com au Authorised Distributor...