10
TEST DIAGNOSTICS
The PROLOGGER has inbuilt diagnostic firmware activated by the protected
command:
Tssss = serial number
Warning: Once you select diagnostic mode, information stored in the logger
may be lost. Always unload the logger before performing test diagnostic
requests and reload the logger after diagnostics.
10.1
Diagnostic
This diagnostic offers several tasks used to test and calibrate the
PROLOGGER during manufacture. Each task corresponds to a single
character sent to the logger via the RS-232 computer interface. After
receiving the Protected Command
Tssss
, the logger enters Diagnostic Mode
and sends the message:
PROLOGGER Diagnostics 7001A V33.0F"
...followed by a menu of diagnostic commands.
The diagnostic then awaits the entry of a Task Command from the host
computer. When a Task Command is received, the logger executes the Task
then awaits another Task Command (unless the previous task exited the
logger from the Diagnostic Mode).
The serial number must be in hex format and entered Least Significant Byte
to Most Significant Byte. For example, the serial number 4001 is hex number
0FA7 and should be entered using the Tssss command in the following order
TA70F.
Warning: Diagnostic Mode is intended for technical maintenance. Certain
protection features are disabled. TAKE CARE.
PROLOGGER Hardware
TEST DIAGNOSTICS 51
Summary of Contents for Starlog Prologger 7001
Page 6: ......
Page 60: ......
Page 64: ......
Page 67: ...02 PROM Verify Fail 03 Mirror Fail 04 Bad checksum in RAM firmware PROLOGGER Hardware 3...
Page 68: ......