© Copyright 2010 TRsystems GmbH , Systembereich Unidor LVCpro-SV13_E.doc Version 1.3 /08.11.10 Page 16 of
20
switches SW202 (channel 1) or SW402 (channel 2) on the basic position 3.
Adjust the maximum output of the measurement object (e.g. the maximum press
force). Continuously up-shift the code-switch SW201 (channel 1) respectively SW401
(channel 2) until reaching the desired output voltage on the analogous output (max.
10V). We recommend a final adjustment of 7V output voltage for maximum load to
still enable an assessment in case of an overload.
If replacing the load amplifier cassette LVCpro-S it is important to maintain the same
adjustment of the code-switches, otherwise there is the risk of overload (falsification
of the measuring results)!
10.
Ordering Data LVCpro-S
Article description
Type
Order-No.:
Load Amplification Cassette, 2-channel
LVCpro-S 02
10 40 4302 0000
Load Amplification Cassette, 2-channel
(low amplification / sensitivity)
LVCpro-S 02-G
10 40 4402 0000
Press Force Sensor black sheath (column
mounting) with 6m* connection cable
JZT 127 / S
50 01 01 01 0000
Press Force Sensor white sheath (con-rod
mounting) with 6m*connection cable
JZT 127 / P
50 01 01 06 0000
Coaxial Cable , open ends on both sides(by the
meter)
MIL-C-17F-RG
058 CU
64200120