43
breakdown/burn
button
9.5
Short Circuit Settings:
In this meter the short circuit current of the meter can also be changed. There are 3
settings available 1.2 mA, 3 mA & 6 mA. Also in case of fault(current goes above short circuit
value) you can select whether to keep running the test or not using the breakdown/burn
settings. To change these settings you can press the icon and then the following
window will be displayed:
You can select the desired short circuit current by clicking on the corresponding button.
And you can change the breakdown/burn setting by clicking on the icon in the bottom left
corner which indicates the present breakdown/burn setting.
Difference between breakdown mode and burn mode:
Breakdown mode:
In this mode the insulation resistance test will be stopped in case the
current in the DUT crosses the selected short circuit current for about 20 seconds. In
breakdown mode the meter displays message. The meter may display one of 2 different
messages after it stops:
⚫
Current limit condition detected
: This is displayed if the voltage at DUT delivered by
meter doesn’t reach the nominal voltage(voltage set by user) and the current crosses the
short circuit current for about 20 sec.