background image

 

ATMOS Operation Manual 

page 7/24 

ATMOS Functions 

The ATMOS device offers various functions for daily laboratory work.  
Apart from the pipette accuracy measurement, it can also support the user with molarity/dilution/serial 
dilution calculations and offers a standard calculator. 
All functions are described below. 

Accuracy Function 

The accuracy function will measure the accuracy of your pipette. 
Click the button “Accuracy” on the main screen. 

Pipette Size Selection Screen 

Select the pipette size you are testing:  

 

Measurement screen

 

After pipette selection or after measurements, you will see the measurement screen. 

Select the nominal 

volume of the 

pipette  

Battery status (it is 

recommended to keep 

ATMOS in charge when 

not in use) 

Go to settings 

screen 

Return to main screen 

Summary of Contents for ATMOS

Page 1: ...ATMOS Operation Manual page 1 24 ATMOS Pipette Accuracy Tester Operation Manual www unipix atmos com sales unipix atmos com...

Page 2: ...ty for errors or missions or for damages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it In no event shall the publis...

Page 3: ...surement screen 7 Measurement Procedure 8 Results Screen 10 Statistics Screen 11 Molarity Function 12 Dilution Function 13 Serial Dilution Function 14 Calculator Function 15 Settings Screen 16 ReadATM...

Page 4: ...n or attempt to repair the instrument without expressed and explicit instructions from UniPix Do not use the instrument in an atmosphere with a danger of explosion Do not allow unauthorized and or unt...

Page 5: ...with liquid in the pipette tip This will permanently damage the instrument no repair is possible Allow the internal temperature to stabilize during at least 30 minutes after unpacking or moving the in...

Page 6: ...tion for your customizable values o Dilution Calculator for dilution of your solution o Serial dilution Calculator for serial dilution by pre defined or customized factors o Calculator Standard calcul...

Page 7: ...ctions are described below Accuracy Function The accuracy function will measure the accuracy of your pipette Click the button Accuracy on the main screen Pipette Size Selection Screen Select the pipet...

Page 8: ...he Volume button allows the selection of sub volumes of the pipette in the case of variable volume pipettes Measurement Procedure 1 Insert the pipette tip firmly in the center port Hold the pipette ve...

Page 9: ...tip inserted in the port Movement should be done during the BLUE phase of progress bar 4 The progress bar at the bottom indicates that the instrument is measuring The pipette and the piston should be...

Page 10: ...rent measurement It will not be stored in the memory NEXT This will start the next measurement The current measurement will be stored in the memory END This will stop the series and display statistics...

Page 11: ...formation on the pipette P and volume pre defined by user ID numeric ID pre defined by user Average volume of the series of n measurements and standard deviation STDEV of the measurement series The Ne...

Page 12: ...numbers a number keyboard is displayed Please note the Enter button is The green frame indicates the value that is calculated To move it and change the value you want to calculate press the button R...

Page 13: ...an insert your numbers Please note the button is Enter The green frame indicates the value that is calculated To change the value you want to calculate press the button R on the lower left side of the...

Page 14: ...ens and you can insert your custom volume Please note the button is Enter The default calculation is a dilution by 3 By clicking on the pink button on the bottom of the display you can select other de...

Page 15: ...r Click on the button Calculator on the main screen to start the tool A standard numbers keyboard shows up and you can calculate Divide X Multiply Subtract Add Result The button AC resets all values T...

Page 16: ...digit will automatically confirm your input Time To change the time click on the button Time You will be asked to type in the time in the following order 1 Hour 2 Minute Please note ATMOS runs on 24h...

Page 17: ...ick on the button Test You will see the following screen P measured pressure Pref reference pressure T1 temperature on sensor 1 T2 temperature on sendor 2 The values of P and Pref should be similar No...

Page 18: ...will approach each other again They should be very similar If this is the case the self test of the device was successful and it runs correctly If the values differ please run the test again maybe wit...

Page 19: ...nder Info Gear symbol on the main screen then Info Firmware V2023 V3 0 and above Verify that your ATMOS device is connected to the computer and turned on Then start ReadATMOS V3 0 on your computer The...

Page 20: ...ce is connected to the computer and turned on Then start ReadATMOS on your computer The following window appears First click the button update ports to identify the port connected to ATMOS Select the...

Page 21: ...file Please select the storage location and enter a file name Then click Save After pushing the Save button the ReadATMOS program will download data from ATMOS The blue bar shows the progress When the...

Page 22: ...P 1000 1000 l 4 l 1 5 l P 200 200 l 0 1 l 0 3 l P 100 100 l 0 5 l 0 15 l P 20 20 l 0 18 l 0 05 l P 10 10 l 0 14 l 0 06 l Figure Atmos accuracy in function of tested volume displacement Technical Speci...

Page 23: ...another pipette Refer to user manual for proper operation Dirty or damaged tip port Clean no liquid or contact your sales person Very bad pipette Verify tip and try another pipette Software shows err...

Page 24: ...of normal wear and tear as well as the failure to follow the instructions of the operating manual UniPix shall not be liable for damage resulting from any actions other than those described in the op...

Reviews: