47
4
SETTINGS AND CALIBRATIONS
47
SETTINGS AND CAL
IBRATIONS
Chapter
4
4-5.
Measurement start condition setting
Set measurement conditions by Measurement Start Condition Setting. The Measurement Start
Condition is as follows:
Operation
Main screen
→
Setting
→
Work Setting
→
Start Setting
1.
Measurement Start Condition
Set the measurement start timing.
Operation
Main screen
→
Setting
→
Work Setting
→
Start Setting
→
Start Condition
①
Only External Input
Measurement starts when the external input signal
“Start” turns OFF
→
ON
→
OFF.
Setting of “2. Measurement Start Level” is disabled.
②
External Input + Load
Measurement starts when the load crosses the
Measurement Start Level after the external input signal
“Start” turns OFF
→
ON
→
OFF.
Load
Measurement
Time or displacement
Start
OFF
ON
Section
Load
Time or displacement
Start
OFF
ON
Start
load
Measurement
Section
Start level
Load
Start point
Start point
×
Not crossing
×
Not crossing
〇
Crossing in the
negative-to-
〇
Crossing in the
positive-to-
positive direction
negative direction