5.SETTING OF FUNCTIONS
52
5-16. Excitation Voltage
This function selects the bridge excitation voltage to be supplied to the strain gauge
sensor.
1) Select setting mode 3.
2) Select excitation voltage.
Set excitation voltage with
and
keys and validate it
with
key.
To return to the indicated value display, press
key.
FNC
SEL
SEL
ENT
Press four times.
DOWN
UP
DOWN
ENT
ESC
Setting Excitation Voltage
Excitation Voltage
1
:
10 V
0
:
2.5V
CAUTION
Use a strain gauge sensor to be connected to the F340A whose
maximum excitation voltage is above the bridge excitation voltage
specified.
If the bridge excitation voltage is greater than the maximum excitation
voltage of the sensor, the sensor may overheat or may be damaged.