24.Setting Items list
161
24. Setting Items list
Mode
Setting item
Initial value
NOV.
RAM
SRAM LOCK
(SW)
LOCK
(SOFT)
Input digits or selections
CH1 CAL.
EXC. Volt
2.5V
◎
◎
○
0: 2.5V 1: 10V
Zero CAL
◎
◎
○
Execution screen
Equiv. CAL. (input value)
2.000
◎
◎
○
4 digits (The explanatory display is changed
according to the mounting condition.)
Equiv. CAL (load)
500.0
◎
◎
○
4 digits
Actual CAL.
500.0
◎
◎
○
4 digits (The explanatory display is changed
according to the mounting condition.)
Digital Flt.
None
◎
◎
0: None 1: 2 2: 4 3: 8 4: 16 5: 32
Analog Flt.
600Hz
◎
◎
0: 10Hz 1: 50Hz 2: 200Hz 3: 600Hz
Decimal Place
xxx.x
◎
◎
○
0: x, 1: x.x 2: x.xx 3: x.xxx
Unit
kN
◎
◎
○
Digital Offset
0.0
◎
◎
◎
4 sign
Increment
0.1
◎
◎
◎
3 digits
CH2 CAL.
Axis
Status display
EXC. Volt
2.5V
◎
◎
○
0: 2.5V 1: 10V (The setting cannot be made
depending on the mounting condition.)
Zero CAL.
◎
◎
○
Execution screen
Equiv.CAL. (input value)
5.000
◎
◎
○
4 digits (The explanatory display is changed
according to the mounting condition.)
Equiv.CAL. (magnitude)
500.0
◎
◎
○
4 digits (The explanatory display is changed
according to the mounting condition.)
Actual CAL.
500.0
◎
◎
○
4 digits (The explanatory display is changed
according to the mounting condition.)
Digital Flt.
None
◎
◎
0: None 1: 2 2: 4 3: 8 4: 16 5: 32
Analog Flt.
600Hz
◎
◎
0: 10Hz 1: 50Hz 2 :200Hz 3: 600Hz
Decimal Place
xxx.x
◎
◎
○
0: x, 1: x.x 2: x.xx 3: x.xxx
Unit
kN
◎
◎
○
Digital Offset
0.0
◎
◎
◎
4 sign (only plus at X-axis setting)
Increment
kN
◎
◎
◎
3 digits
Operation Mode
Operation Mode
Multi-hold
◎
◎
0: Multi-hold 1: Wave Comp 2: Hysteresis
3: Wave & Displace 4: Hysteresis 2
Sample Rate
1kHz
◎
◎
0: 4k 1: 2k 2: 1k 3: 500 4: 200 5: 100
Code Set
[0-31] ch
Ch. No.
0
◎
2 digits. 0 will result when the power is turned on.
Parameter Copy
0
◎
◎
2 digits
Select CH
CH1
◎
◎
0: CH1 1: CH2
Hold Set
None
◎
◎
HI Limit
+100.0
◎
◎
4 sign
HI-HI Limit
+999.9
◎
◎
4 sign
LO Limit
+1.0
◎
◎
4 sign
LO-LO Limit
-999.9
◎
◎
4 sign
DPM Hi
999.9
◎
◎
4 digits
DPM LO
0.0
◎
◎
4 digits
Wave Start Level
+1.0
◎
◎
4 sign
Wave End Level
+0.0
◎
◎
4 sign
Hold Timer
1000
◎
◎
4 digits
P&V Difference
1.0
◎
◎
4 digits
P&V Judg. Rate
25%
◎
◎
0: 25% 1: 50% 2: 75% 3: 125% 4: 150% 5: 200%
6 :300% 7: 400%
P&V Ordinal
1
◎
◎
1 digits
Minimum Slope
1.0
◎
◎
4 digits
Interval A
20
◎
◎
3 digits
Interval B
20
◎
◎
3 digits
Pre-Point
0
◎
◎
2 digits
Inhibit Timer
0
◎
◎
4 digits
Hys. Interval
1.0
◎
◎
4 digits
F-T Limit
4000
◎
◎
4 digits
F-T Output
ON
◎
◎
0: ON 1: OFF
HH/LL Mode
Hold
◎
◎
0: Hold 1: Sample
Lv. Axis Select
Y Axis
◎
◎
0: Y Axis 1: X Axis
Hold Start Level
-9999
(
Invalidity
)
◎
◎
4 sign
Auto Code Up
OFF
◎
◎
0: ON 1: OFF
Code Up Point
0
◎
◎
4 digits
HI-LO Relative
OFF
◎
◎
0: ON 1: OFF
DPM HI-LO Mode
Peak value
◎
◎
0: Hold Value 1: Peak Value 2: Both
DPM Hold HI
9999
◎
◎
4 digits
DPM Hold LO
0
◎
◎
4 digits
U. Area Output
ON
◎
◎
0: ON 1: OFF
0:kg
1:t
2:g
3:N
4:kN
5:MN
6:N•m
7:Pa
8:kPa
9:MPa
10:bar
11:N/m
2
12:mm
13:cm
14:
μ
m
15:rad
16:deg
17:lb
18:oz 19:dyne
20:psig
21:ftlb
22:inlb
23:inoz
24:ft
25:in
26:None
0:kg
1:t
2:g
3:N
4:kN
5:MN
6:N•m
7:Pa
8:kPa
9:MPa
10:bar
11:N/m
2
12:mm
13:cm
14:
μ
m
15:rad
16:deg
17:lb
18:oz 19:dyne
20:psig
21:ftlb
22:inlb
23:inoz
24:ft
25:in
26:None