25.Statement of Conformation to EC Directives
163
25.Statement of Conformation to EC Directives
Unipulse F395 Weighing Controller conforms to The EC Directives (based on Council of
the European Communities, and is allowed to affix CE mark on it.
* Lower Voltage Directives EN61010-1
* EMC Directives EN61326-1
The following notice must be observed when you install F395 unit.
1. Since F395 is defined as an open type (unit to be fixed or built-in), it must be
fixed or bolted to frame or solid board securely.
2.
The power cable attached to this product as standard equipment can be used in the
AC100V power supply in Japan. (Official ratings voltage AC125V)
Please use the power cable authorized in the country when you use this product
outside Japan.
3. Use shielded cable for connection with components other than power supply,
such as sensor, input/output signals and optional device.
4. Attach separate type ferrite core (supplied as standard item with the unit) on
terminal box side of power supply cable, cable for power supply output and
also on connecting cable to sensor.
Note:
・
Lightning surge protector is not included in standard supply for F805.
・
For installation of lightening surge protector, see next page.
EN55011
,
EN61000-4-2
,
EN61000-4-3
,
EN61000-4-4
EN61000-4-5
,
EN61000-4-6
,
EN61000-4-8
EN61000-4-11
,
EN61000-3-2
,
EN61000-3-3
Notice
・
Combined use of F395 and lightning surge protector conforms to EN61000-4-5
of EMC directives. (lightning surge immunity)
・
CC-Link option do not conform to EC Directives.