All movement between tabs and menu 
options is performed by pressing 
the select and arrow keys next to the 
display. 
•  To select a tab or menu point, press 
Select.  
•  To adjust a menu point, press the up 
and down arrow keys. 
•  To exit a menu point and go back to 
the menu, press the left arrow key. 
 

a

djusting

 

the

 

pre

-

aMplificatiOn

You may need to adjust the pre-amplification range of the sensor. This is done in the Measurement 
tab on the Microsensor Multimeter by selecting the relevant sensor. The appropriate range is the 
lowest range that can accommodate the highest signal to be measured. The range 1 mV/pA is 
selected per default. If the expected 
maximum signal in this range exceeds 
4500 mV, the range should be changed 
to 0.1 mV/pA. If the expected maximum 
signal is less than 500 mV in the 1 mV/pA 
range, then the 10 mV/pA range should 
be selected. 

c

Ompatibility

s

ensOrs

The Microsensor Multimeter can be used 
only with sensors with LEMO connec-
tions. Adaptors are available for sensors with BNC connections. 
In the standard configuration, the Microsensor Multimeter has amperometric channels on channel 1 
and 2, a potentiometric channel on channel 3, and a temperature channel on channel 4. For custom 
configured instruments, please refer to your order confirmation for channel setup. 

Amperometric sensors: O

2

, H

2

, H

2

S, N

2

O, NO, NO

2

-

, NO

3

-

, Flow, Diffusivity, APOX, TPOX, STOX, 

Potentiometric sensors: pH, Redox 
Temperature

Sensors that are automatically recognized and polarized: O

2

, H

2

S, pH, Redox, Temp.

All other sensors AND all BNC sensors connected via an adapter are recognized as “Sensor for manual 
setting” and polarization is set manually. Please refer to the relevant sensor manual for polarization 
values. 