GENERAL INFORMATION

Thank you for choosing the Unisense SulfiLogger™ sensor - the only H

2

S sensor on the market enabling

direct measurement of hydrogen sulfide concentrations in either liquid or gas phase, easy plug’n’play
installation in harsh environments such as sewer networks, and easy maintenance.

S

AFETY

Please read this manual carefully before unpacking, installing or operating this equipment. Failure to do so
may result in serious injury to operators or other persons, or damage to equipment. Do not use or install
the SulfiLogger™ sensor and auxiliary equipment in any other way than specified in this manual. Special
attention must be paid to all danger and caution statements. Do not use the SulfiLogger™ sensor outside
the specified electrical, mechanical and thermal parameters as specified. During the installation, please

wear personal protective equipment (PPE) according to the safety regulation of the installation site.

* For SulfiLogger™ X1 series sensors only (EX):

Pay close attention to other EX equipment used in conjunction with the SulfiLogger™ sensor.

Observe the specifications in the certificate as well as national and local regulations.

Ensure that all personnel are suitably qualified.

Always use approved EX barriers for installation.

Chemical and gas handling safety

Operators handling the SulfiLogger™ sensor in potentially dangerous environments should be trained in the
environmental safety and need of PPE for the given application.

DANGER

Hydrogen sulfide is a highly toxic gas and special precautions are needed to avoid the risk of injury
or death. Users of this product are advised to familiarize themselves with safety procedures, use of
personal protective equipment (PPE) and the correct use of gases and/or chemicals, and to carefully
read all relevant Safety Data Sheets.

S

ENSOR HANDLING

Pay special attention to always orient the SulfiLogger™ sensor with the sensor head facing downwards,
including when handling, transporting, installing, storing and calibrating the sensor.

!

CAUTION

Orienting the SulfiLogger™ sensor with the sensor head pointing upwards may damage the sensor.

2

Summary of Contents for SULFILOGGER

Page 1: ...SulfiLogger_Manual_EN_203 pdf SulfiLogger User Manual COPYRIGHT 2019 UNISENSE A S VALID FOR SERIAL NUMBERS FROM S N 1004000 REV 2019 11 19...

Page 2: ...leaning 4 Calibration 4 Operating principles 5 Electrochemical measurements of hydrogen sulfide 5 Measuring in different media unit conversions 5 Specifications 6 SulfiLogger sensors 6 Technical speci...

Page 3: ...ut of the use the results of use or the inability to use this product Unisense A S assumes no liability for damages on instruments designed for indoor use due to unintended field use or exposure to du...

Page 4: ...ulfiLogger X1 series sensors only EX Pay close attention to other EX equipment used in conjunction with the SulfiLogger sensor Observe the specifications in the certificate as well as national and loc...

Page 5: ...fiLogger X1 and standard versions of the sensor are labeled SulfiLogger S1 If your product is an ATEX certified product X1 series sensors it is very important that you carefully follow the instruction...

Page 6: ...lled using a flow cell please detach the sensor from the flow cell before starting the calibration If needed replace the O ring on the sensor before re installation 1 2 3 4 5 6 SUCCESS FAILURE 6 1000...

Page 7: ...of the daily variations in ventilation conditions in the head space of the sewer and it is independent of the exact mounting position of the sensor in the liquid the sulfide concentration is fairly ho...

Page 8: ...t 1 of full range Operating humidity range 0 100 Design temperature range 20 C to 60 C 4 F to 140 F Operating temperature range 0 C to 40 C 32 F to 104 F MECHANICAL SPECIFICATIONS Length 240 mm Diamet...

Page 9: ...lose to the end of pipe of the force main as possible to get a realistic measure of the initial H2S concentration The SulfiLogger S1 X1 1120 5mgL sensor can be mounted at access points directly on the...

Page 10: ...hould be stated on the bottle or can be provided by the vendor If judging the accuracy of the SulfiLogger sensor based on liquid phase measurements there are various ways to do this all of which are p...

Page 11: ...Logger sensor is temperature compensated the accuracy of the sensor is highest if the calibration is done at a temperature close to the temperature of subsequent measurements Problem During a calibrat...

Page 12: ...highly toxic gas and special precautions are needed to avoid the risk of injury or death Furthermore the installa tion of the SulfiLogger sensor in applica tions such as sewer networks or other facil...

Page 13: ...he Sul fiLogger sensor and ensure that the pressure pipe system does not exceed the pressure limit Bypass flow using a flow cell A portion of the flow can be di rected bypassed into a flow cell on whi...

Page 14: ...SulfiLogger_Manual_EN_203 pdf UNISENSE UNISENSE DENMARK www unisense com info unisense com...

Reviews: