MAINTENANCE

C

LEANING

SulfiLogger™ sensors installed in wastewater environments may require regular cleaning depending on the
fouling conditions. To clean the sensor, remove any accumulated rags and gently wipe the sensor head with
a cloth. Use water or a mild soap if needed. For more information, please see page 7.

C

ALIBRATION

The SulfiLogger™ sensor is factory calibrated and is as such ready to use right out of the box. It is recom-

mended to calibrate the sensor before installation and then subsequently once every month to maintain
accuracy. The calibration should be performed at your measurement location and takes approx. 5 - 10
minutes. To calibrate the sensor you must use a SulfiLogger™ calibration kit consisting of a calibration gas
bottle, a flow regulator and a calibration cap. During calibration, you should avoid orienting the sensor with
the sensor tip pointing upwards. If the sensor is installed using a flow cell, please detach the sensor from the
flow cell before starting the calibration. If needed replace the O-ring on the sensor before re-installation.

1

2

3

4

5

6

SUCCESS

FAILURE

6

1000 

ppm

Cal. gas: 1000 ppm

Retry or contact support

WAIT...

RED/GREEN

GREEN

RED

FULLY 

OPEN

min.

5 bar

CLOSE

1. Verify that the pressure gauge on the flow regulator reads at least 5 bar. Make sure that the range

and concentration on the calibration gas bottle and on the sensor is the same.

2. Attach the calibration cap to the sensor head and make sure to align the arrow on the calibration cap

with the position of the sensor’s indicator light. The sensor has successfully entered its calibration

mode when the indicator light switches to a pattern of alternating red and green flashes.

3. Fully open the regulator valve and wait for the calibration to complete.

4. When the calibration has completed, the indicator light will change to a pattern of continuous green

flashes if the calibration has succeeded or continuous red flashes if it has failed. In either case, close
the regulator valve (5) and remove the calibration cap from the sensor (6).

4

Summary of Contents for SULFILOGGER

Page 1: ...SulfiLogger_Manual_EN_203 pdf SulfiLogger User Manual COPYRIGHT 2019 UNISENSE A S VALID FOR SERIAL NUMBERS FROM S N 1004000 REV 2019 11 19...

Page 2: ...leaning 4 Calibration 4 Operating principles 5 Electrochemical measurements of hydrogen sulfide 5 Measuring in different media unit conversions 5 Specifications 6 SulfiLogger sensors 6 Technical speci...

Page 3: ...ut of the use the results of use or the inability to use this product Unisense A S assumes no liability for damages on instruments designed for indoor use due to unintended field use or exposure to du...

Page 4: ...ulfiLogger X1 series sensors only EX Pay close attention to other EX equipment used in conjunction with the SulfiLogger sensor Observe the specifications in the certificate as well as national and loc...

Page 5: ...fiLogger X1 and standard versions of the sensor are labeled SulfiLogger S1 If your product is an ATEX certified product X1 series sensors it is very important that you carefully follow the instruction...

Page 6: ...lled using a flow cell please detach the sensor from the flow cell before starting the calibration If needed replace the O ring on the sensor before re installation 1 2 3 4 5 6 SUCCESS FAILURE 6 1000...

Page 7: ...of the daily variations in ventilation conditions in the head space of the sewer and it is independent of the exact mounting position of the sensor in the liquid the sulfide concentration is fairly ho...

Page 8: ...t 1 of full range Operating humidity range 0 100 Design temperature range 20 C to 60 C 4 F to 140 F Operating temperature range 0 C to 40 C 32 F to 104 F MECHANICAL SPECIFICATIONS Length 240 mm Diamet...

Page 9: ...lose to the end of pipe of the force main as possible to get a realistic measure of the initial H2S concentration The SulfiLogger S1 X1 1120 5mgL sensor can be mounted at access points directly on the...

Page 10: ...hould be stated on the bottle or can be provided by the vendor If judging the accuracy of the SulfiLogger sensor based on liquid phase measurements there are various ways to do this all of which are p...

Page 11: ...Logger sensor is temperature compensated the accuracy of the sensor is highest if the calibration is done at a temperature close to the temperature of subsequent measurements Problem During a calibrat...

Page 12: ...highly toxic gas and special precautions are needed to avoid the risk of injury or death Furthermore the installa tion of the SulfiLogger sensor in applica tions such as sewer networks or other facil...

Page 13: ...he Sul fiLogger sensor and ensure that the pressure pipe system does not exceed the pressure limit Bypass flow using a flow cell A portion of the flow can be di rected bypassed into a flow cell on whi...

Page 14: ...SulfiLogger_Manual_EN_203 pdf UNISENSE UNISENSE DENMARK www unisense com info unisense com...

Reviews: