E.  If "single supply" operation is desired, this allows the power supply to be used as twice the voltage and 

rating current simply by using the negative (black) terminal of    the SLAVE supply and the positive (red) 
terminal of    the MASTER supply, the configuration as shown in Fig.5-3. 

 

Fig.5-3 Single Supply 

 

If the chassis or common of the equipment being powered is separate from both positive and negative 
polarity power inputs. The output of the SLAVE (negative) supply is tracking the output of the MASTER 
(positive) supply. The configuration is shown in Fig.5-4. 

 

Fig.5-4 Positive and Negative Supply 

 

5-4-3. Parallel Tracking Operation 

In the parallel tracking mode of operation, both supplies are strapped together (in parallel). This allows for a 
rating voltage supply with a double rating current capability. Only the MASTER output terminals are used for 
parallel tracking operation. In the parallel tracking mode, the SLAVE supply output voltage and current are 
tracking the MASTER supply output voltage and current. 
A.  Set the power supplies to the TRACKING PARALLEL mode by engaging both TRACKING switches. 
B.  Because both voltage and current of the SLAVE supply track the MASTER supply, the maximum current 

and voltage are set using the MASTER controls. Using the MASTER supply output jacks, follow the 
instructions for "Setting Current Limit" (5-2 Section). Remember that the actual current output at the 
MATER supply output jack is double the reading on the SLAVE indicator meter.. Then push the output 
on/off key to enter voltage and current setting. 

C.  Adjust the output voltage to the desired level using the MASTER VOLTAGE controls. 
D.  Turn off the power supply and the equipment to be powered during hook-up. 
E.  Connect the positive polarity of the device being powered to the red (+) terminal of the MASTER power 

supply. 
 
 
 

Summary of Contents for PS-303DU

Page 1: ...Multi channel DC Power Supply PS 303DU 305DU 3010DU 6005DU PS 2303 2305 Operation Manual V1 0 BENCHTOP INSTRUMENT...

Page 2: ...ear Panel Control 8 5 OPERATION INSTRUCTIONS 9 5 1 Precaution 9 5 2 Setting Current Limit 9 5 3 Constant Voltage Constant Current Characteristic 9 5 4 Operation Mode 10 5 4 1 Independent Operation 10...

Page 3: ...at the maximum output current or can be continuously adjusted throughout the output range The front panel current controls can be used to establish the output current limit overload or shout circuit...

Page 4: ...05DU 0 60V x2 0 5A x2 Built in 16 Dimensions PS 3010DU 6005DU 255Wx155Hx460D mm PS 303DU 305DU 2303 2305 255Wx155Hx330D mm 3 SPECIFICATIONS 3 1 Operation Mode 1 Independent Two independent outputs and...

Page 5: ...0mV Fixed output Output voltage 5V 0 25V Output current 3A Load regulation 25mV 110 220VAC rated input Ripple noise 2mVrms 5Hz 1MHz Display Voltmeter 3 digits 0 4 LED display Ammeter 3 digits 0 4 LED...

Page 6: ...4 Fig 4 2 b PS 2303 2305 Front panel Fig 4 2 c PS 3010DU 6005DU Front panel...

Page 7: ...5 Fig 4 3 d PS 303DU 305DU Rear panel Fig 4 4 e PS 2303 2305 Rear panel...

Page 8: ...6 Fig 4 5 f PS 3010DU 6005DU Rear panel Fig 4 6...

Page 9: ...when the MASTER supply is in the constant current operation 13 C C indicator Lights when the SLAVE supply is in the constant current operation Also lights when the TRACKING PARALLEL mode is selected 1...

Page 10: ...tput terminal for 5V supply 24 Output terminal Negative polarity output terminal for 5V supply 29 Output indicator Lights when switch is engaged in 30 Output ON OFF switch DC power supply output when...

Page 11: ...efore adding load adjust the voltage knob to the minimum value then turn on the instrument Next turn the voltage current adjustment knob to set targeted values 3 Output voltage overshoot Voltage betwe...

Page 12: ...ry will place a heavy load on the supply and it will operate in the constant current mode which may be adjusted for a 1 amp charging rate As the battery becomes charged and its voltage approaches 13 8...

Page 13: ...double the displayed value For example if the MASTER display is set for voltage metering and the SLAVE display for current metering the output voltage across the MASTER positive red terminal and the S...

Page 14: ...nly the MASTER output terminals are used for parallel tracking operation In the parallel tracking mode the SLAVE supply output voltage and current are tracking the MASTER supply output voltage and cur...

Page 15: ...correct this situation the load on the supply must be decreased so that no more than 3 amps of current are sucked 6 MAINTENANCE WARNING The following instructions are for use by qualified personnel on...

Page 16: ...Make sure the power cord is unplugged 2 Change the AC select switch to the desired line voltage position 3 A change in line voltage may also require a corresponding change of fuse value Install the c...

Reviews: