CHAPTER 6: CALIBRATION
HF25D DC RESISTANCE WELDING SYSTEM
990-333
6-2
Calibration Procedure
Initial Calibration Setup
1.
Connect the calibration setup to the Control as shown.
2.
Turn the Control ON.
3.
Press the
MENU
key to bring up the
main menu screen.
<MAIN MENU>
1.
SETUP
5.
COMMUNICATION
2. WELD COUNTER
6. RELAY
3. COPY A SCHEDULE
7. CALIBRATION
4. SYSTEM SECURITY
8. RESET TO
DEFAULTS
4.
Press the
7
key to select the calibration
function. This will bring up the
calibration caution screen.
<CALIBRATION CAUTION>
CALIBRATION SHOULD BE PERFORMED BY A
QUALIFIED TECHNICIAN ONLY.
REFER TO MANUAL FOR CALIBRATION SETUP.
▼
Next, Menu menu
5.
From the calibration caution screen,
press to continue. This will bring up
the pre-calibration screen.
6.
From the pre-calibration screen, press
the
2
key to start the sequence of on-
screen calibration instructions.
<PRE-CALIBRATION>
1. TEST HF25
2. CALIBRATE HF25
3. RESET CALIBRATION
4. SET CURRENT SHUNT VALUE
5. TEST SERIAL PORT
▼
Next, Menu menu