5

Safety Measures

In order to avoid instrument damage, it is advi-
sed to remove batteries when not using the in-
strument over a certain period of time. Howe-
ver, should the instrument be contaminated by
leaking battery cells, you are kindly requested to
return it to the factory for cleaning and inspec-
tion.

Instruments must be stored in dry and closed areas.
In the case of an instrument being transported in ex-
treme temperatures, a recovery time of minimum 2
hours is required prior to instrument operation. 

3.0 Safety Measures

The UNITEST TELARIS LOOP has been designed
and checked in accordance with the safety regulati-
ons for Electronic Test and Measurement Instru-
ments EN 61010 and IEC 1010, and left our factory
in a safe and perfect condition. To maintain this con-
dition the user must pay attention to the safety re-
ferences contained in this instruction manual.

The respective accident prevention regulations
established by the professional associations re-
garding electrical systems and equipment must
be strictly met for all tasks.

In order to avoid electrical shock, the valid sa-
fety and VDE regulations regarding excessive
contact voltages must receive the utmost atten-
tion, when working with voltages exceeding
120V (60V) DC or 50V (25V)rms AC. The values
in brackets are valid for limited ranges (eg for
medicine and agriculture).

If the operator’s safety can no longer be gua-
ranteed, the instrument is to be put out of ser-
vice and protected against use. The safety is no
longer guaranteed, if the instrument:

• shows obvious damage
•  does not carry out the desired measurement.
•  has been stored for too long under unfavou-

rable conditions.

• has been subjected to mechanical stress du-

ring transport.

The instrument may only be used within the
operating ranges as specified in the technical
data section.

Avoid any heating up of the instrument by direct
sunlight to ensure perfect functioning and long
instrument life.

The Earth Loop Resistance Measurement In-
strument complies with the overvoltage class
CAT III / 300V and may only be used for mains
systems of maximum 230 V (fused at maximum
16 A).

Only use the mains cable supplied.  The unit
was calibrated with this cable and only in this
way can the resistance values measured be
exactly indicated.

Summary of Contents for TELARIS LOOP

Page 1: ...TELARIS LOOP UNITEST Instruction Manual Cat No 8978...

Page 2: ...2...

Page 3: ...ement 8 6 2 Internal Mains Resistance Measurement 9 6 3 RPE RN RL Measurement 10 6 4 AC Voltage Measurement VAC 11 6 5 Frequency Measurement f 11 6 6 Contact Electrode 11 7 0 Display Indications 12 8...

Page 4: ...ST TELARIS LOOP is charac terised by the following features A unit designed to allow the user to quickly esta blish earth loop impedance mains internal resi stance and prospective short circuit curren...

Page 5: ...or all tasks In order to avoid electrical shock the valid sa fety and VDE regulations regarding excessive contact voltages must receive the utmost atten tion when working with voltages exceeding 120V...

Page 6: ...s well as for trouble shooting Simultaneously to measurement the instrument TELARIS LOOP determines by calculation the short circuitcurrent Theshort circuitcurrentmust be known to guarantee that an ex...

Page 7: ...for mains lead Test lead 14 LCD 15 Key Vc UL 50V 25V 25V for medical and agricul tural uses 16 Contact electrode 17 Measurement function switch off ZL IPSC Loop impedance short circuit current RI IPSC...

Page 8: ...e mains lead to the mains connector 13 Select the measurement function R via the fun ction switch 17 Connect the test cables mains plug to UUT socket by respecting correct polarity Set a voltage conta...

Page 9: ...ement results switch off all loads or disconnect them from the mains Connect the mains lead to the mains connector 13 Select measurement function via the function switch 17 Select contact voltage limi...

Page 10: ...ressing the key Display 23 you may now select between the different measure ment variables If a value exceeds the measurement limit the Attention symbol appears If this result is sel ected with the ke...

Page 11: ...t the measure ment instrument from UUT Also refer to chapter 7 0 No possibility to store the measurement re sults 6 5 Frequency Measurement Never apply tvoltages exceeding 300V to input sockets Connec...

Page 12: ...ceeded Contact voltage limit 50 V Resistance measurement Zl Ipsc ZL 2 88 Contact voltage limit 50 V Resistance measurement Zl Ipsc RL excessive 19 99 Measurement is interrupted Contact voltage limit 5...

Page 13: ...surement result by pressing key Store 19 Upon completion of storing a short signal is au dible and the symbol M 11 appears in the display for memory location In case of a full memory the symbol M blin...

Page 14: ...rom the OFF position to the de sired function If the Battery is almost dead the symbol 8 appears in the indication The piece of equipment works furthermore however without claim on retention of the sp...

Page 15: ...tery cells Never throw a battery into a fire as this could lead to an explosion Never expose batteries to humi dity Please consider your environment when you dis pose of your used batteries They belon...

Page 16: ...01 Ohm 20 199 9 Ohm 0 1 Ohm 5 rdg 5 Digit 200 Ohm 1 999 kOhm 1 Ohm Internal Resistance RI Range Resolution Tolerance 0 05 19 99 Ohm 0 01 Ohm 5 rdg 5 Digit Resistance Neutral Conductor RN Range Resolu...

Page 17: ...z 0 1 Hz 1 rdg 2 Digit 400V RI 1 M Contact voltage detection UL 25V 25V 5 15 UL 50V 50V 5 15 For large mains voltage fluctuations during measurements the measurement error may increase Auto power off...

Page 18: ...18...

Page 19: ...oumis un contr le technique permanent Ces m mes quipements de contr le sont calibr s r guli rement l aide d appareils de r f rence calibr s selon les di rectives et normes en vigueur dans les laborato...

Page 20: ...e with unda maged warranty label Any damage due to dropping or incorrect handling are not covered by the war ranty If the instrument shows failure following ex piration of warranty our service depart...

Reviews: