- If you experience any interference, do
not use the hearing aids and contact the
manufacturer of the active implant. Please
note that interference can also be caused
by power lines, electrostatic discharge,
airport metal detectors etc.
Special medical or dental examination
including radiation described below, may
adversely affect the correct functioning of
your hearing aids. Remove and keep them
outside the examination room/area before
undergoing:
- Medical or dental examination with X-ray
(also CT scan).
- Medical examinations with MRI/NMRI
scans, generating magnetic fields.
- Hearing aids don’t need to be removed
when passing security gates (airports etc.).
If X-ray is used at all, it will be in very low
doses, and will not affect the hearing aids.
Only use the charger and power supply
provided by the manufacturer for charging.
Otherwise, the device(s) may be damaged.
Use of accessories, transducers and cables
other than those specified or provided by the
manufacturer of this equipment could result
in increased electromagnetic emissions or
decreased electromagnetic immunity of this
equipment and result in improper operation.
Your hearing aids operate in the frequency
range of 2.4 GHz–2.48 GHz. When flying
please check if flight operator requires
devices to be switched into flight mode.
Portable RF communications equipment
(including peripherals such as antenna
cables and external antennas) should be
used no closer than 30 cm (12 inches) to any
part of the hearing aids, including cables
specified by the manufacturer. Otherwise,
the equipment’s performance could be
degraded.
HelloGO Ready-To-Wear hearing devices
should only be used with the designated
earpiece and receiver.
The USB port of the charger is to be used for
the described purpose only.