ASSEMBLY
The following photos show how to assemble the various modules
in the order of assembly.
When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching
any parts or touching glass surfaces.
Installing the Focus Arm
For microscopes ordered with the Plain Focusing or LED stands the focus mount should be positioned at
the higher position on the track. There are two positions on the dovetail track in which the focus mount
can be installed. The positions are spaced 55mm apart. Should you need to lower the position of the
focus mount for the future please follow the following steps:
1. Remove the viewing head, optical body and objective from the focus mount by loosening
the locking screw on the focus mount.
2. Using the allen key that was supplied with your stand loosen the allen screw at the rear of
the focus mount that connects the focus mount to the dovetail.
3. Align the screw hole opening on to the lower screw opening on the dovetail.
4. Align the two pins on the focus arm into the grooves on the dovetail slider.
5. Tighten the allen screw into place.
6. Reinsert the viewing head, optical body and objective into the focus mount and lock the
assembly back into the focus mount by tightening the locking screw.