6.- EQUIPMENT MAINTENANCE
- 78 -
6.4.- Replacing fuse
ATTENTION: Before changing the fuse, ALWAYS unplug the
power cable. Always use fuses of the value and type specified
(see
chapter 9: Specifications)
.
In order to avoid possible supply voltage peaks that might damage the UM2B, the
unit incorporates a fuse, housed in the lower part of the mains connector (5). The fuse-
holder may be slid out by hand or using a small screwdriver. Inside (inner) is the operating
fuse and, in an adjacent compartment (outer), a spare.
The very arrangement of the fuse-holder makes it necessary to disconnect
the power cable. Nevertheless, great care should be taken to ensure
safety, and the cables should be disconnected from the equipment being
tested. Also, the UM2B should be moved away from this equipment.
After having carried out the aforementioned safety processes, replace the fuse as
follows:
1
Turn off the switch on the rear panel
2
Disconnect the cables from the equipment being tested.
3
Disconnect the power cable.
4
Open the fuse-holder.
5
Remove the blown fuse (the inner one).
6
Insert the adequate spare fuse.
7
Close the fuse-holder.
Summary of Contents for UM2B
Page 2: ......
Page 67: ...5 SOFTWARE DESCRIPTION 65 Figure 5 46 a First page of the report Recovery Voltage ...
Page 68: ...5 SOFTWARE DESCRIPTION 66 Figure 5 47 a Second page of report Recovery Voltage ...
Page 69: ...5 SOFTWARE DESCRIPTION 67 Figure 5 48 Third page of report Recovery Voltage ...
Page 70: ...5 SOFTWARE DESCRIPTION 68 Figure 5 46 b First page of the report Insulation Resistance ...
Page 71: ...5 SOFTWARE DESCRIPTION 69 Figure 5 47 b Second page of the report Insulation Resistance ...
Page 87: ...8 TECHNICAL SUPPORT 85 Description of fault Materials returned on of of the year Table 8 1 ...
Page 95: ...APPENDIX B CONTROL SOFTWARE 93 APPENDIX A CE CONFORMITY DECLARATION ...