Maintenance
Before performing any maintenance on the 500 Series Thermoelectric Gas Cooler, ensure that all plant safety
procedures are followed.
The 500 Series Cooler is designed for maintenance free operation, but if any is required, ensure power has been removed
before maintenance is performed.
For the best performance of the cooler, the following maintenance schedule is recommended:
Maintenance Activity
Frequency
Peristaltic pump
Replace tubing every 3 months
Diaphragm sample pump
Replace diaphragm every 6 months
Clean heat exchanger
Annually
Inspect heat sink fins
Monthly