Page 6 of 25
Page 7 of 25
MAN.600.REVF.07012015
Specifications
OPERATING SPECIFICATIONS
Sample Flow Rate
0 to 5 l/m
Maximum Inlet Temperature
Stainless Steel Heat Exchanger
700
°
F (370
°
C)
Kynar/Glass Heat Exchanger
300
°
F (149
°
C)
Maximum Inlet Gas Dew Point
180
°
F (82
°
C)*
Maximum Inlet Water Concentration
50%*
Minimum Ambient Temperature
32
°
F (0
°
C)
Maximum Ambient Temperature
105
°
F (41
°
C)*
Maximum Cooling Power
126 BTUs per hour (120 kJ/hr)
Outlet Sample Dew Point
39
°
F (4
°
C), adjustable
Maximum Input Power (Model/Wattage)
622/78 watts 632/78 watts 642/150 watts
624/108 watts 634/108 watts 644/150 watts
Voltage
12VDC (Model 6X2) or 24VDC (Model 6X4)
Electrical Classification
General Purpose/Safe Area
Dimensions
9 1/4" H x 8 7/8" W x 6 3/8" D
Weight
18 lbs (8kg)
Soluble Gas Removal Rates
NO
0% loss
NO
2
<10% loss
SO
2
< 2% loss
CO
0% loss
CO
2
< 2% loss
* AT REDUCED FLOW RATES, SEE CAPACITY CHART.