Page 14 of 43
Page 15 of 43
MAN.SCPSCU.REVD.03022015
Maintenance
Before performing any maintenance on the cooler, ensure that all plant safety procedures are followed. As with any
electrical device, ensure power is removed before performing any procedures.
The cooler is designed for maintenance free operation but if any is required, ensure power has been removed before
maintenance or repair is performed.
For the best performance of the cooler, the following maintenance schedule is recommended:
Maintenance Activity
Frequency
Peristaltic pump
Replace tubing every 3 months
Diaphragm sample pump
Replace diaphragm every 6 months
Clean heat exchanger
Annually
Inspect heat sink fins
Monthly
MAINTENANCE PROCEDURES
REPLACEMENT OF PERISTALTIC TUBING (IF EQUIPPED)
1.
Please refer to manufacturers website for instructions: bit.ly/1zfmrzt
2.
REPLACEMENT OF SAMPLE PUMP DIAPHRAGM
1.
Please refer to manufactures website for instructions:
INSTALLING OR REPLACING HEAT EXCHANGERS
REMOVING THE HEAT EXCHANGER
1.
Remove the inlet and outlet tubes by loosening the compression fittings. Always use a backup wrench on the fitting
body to ensure no damage to the heat exchanger occurs.
2.
Remove the drain fitting using the same procedure as the inlet/ outlet. Remove the drain fittings from the exchanger.
Use a backup wrench on the lower heat exchanger hex to prevent damage to the exchanger.
REPLACING THE HEAT EXCHANGER
1.
Dry and clean the heat exchanger opening in the heat transfer block using a dry, lint-free cloth (If reusing the heat
exchanger, clean the outside as well.) Dried heat transfer paste can be removed by using a very fine abrasive pad
wrapped around a drill bit.
2. Smear the outer diameter of the heat exchanger with heat transfer paste.
3.
Gently push the heat exchanger into the heat transfer block until the head is fully seated against the insulation on top.
4.
Reinstall the drain fitting. Ensure pipe tape is used on the pipe threads before installation. Use a backup wrench on
the heat exchanger lower hex to prevent damage to the exchanger.
5.
Reconnect the drain, inlet and outlet tubes.