Page 4 of 43
Page 5 of 43
MAN.SCPSCU.REVD.03022015
Specifications
OPERATING SPECIFICATIONS
Sample Flow Rate
0 to 8 l/m **
Maximum Inlet Temperature
Stainless Steel Heat Exchanger
700
°
F (370
°
C)
Kynar/Glass Heat Exchanger
280
°
F (138
°
C)
Maximum Inlet Gas Dew Point
194
°
F (90
°
C)*
Maximum Inlet Water Vapor Content
70%*
Minimum Ambient Temperature
34
°
F (0
°
C)
Maximum Ambient Temperature
105
°
F (41
°
C)*
Maximum Cooling Power (Second Stage)
252 BTUs per hour (240 kJ/Hr)
Outlet Sample Dew Point
39
°
F (4
°
C)
Maximum Input Power
75-700 Watts***
Voltage
95-125VAC 50/60 Hz or
190-250VAC 50/60 Hz
(External fuse required or 20A or less)
Electrical Classification
General Purpose
Temperature Classification
FM/CSA T3A, ATEX 143
°C (T3)
SCP Dimensions
38 1/2" H x 20 3/4" W
SCU Dimensions
31" H x 19" W x 12" D
Weight
75 lbs (34kg)
Soluble Gas Removal Rates
NO
0% loss
NO
2
<10% loss
SO
2
< 2% loss
CO
0% loss
CO
2
< 2% loss
*AT REDUCED FLOW RATE ABOVE 77°F. (25°C.) AMBIENT.
**SEE GAS COOLER CAPACITY CHART BELOW.
***WATTAGE DEPENDANT ON SELECTED COOLER. CONTACT FACTORY FOR SELECTED COOLER WATTAGE.
3000 SERIES - COOLER CAPACITY DATA
Ambient 77°F/25°C
Water Vapor
Ambient 90°F/25°C
Water Vapor
Ambient 105°F/25°C
Water Vapor
12%
15%
30%
50%
12%
15%
30%
50%
12%
15%
30%
50%
3040
5 l/m
4 l/m
2.8 l/m
1.4 l/m
5 l/m
4 l/m
2 l/m
1 l/m
2.8 l/m
2.4 l/m
1.2 l/m
0.6 lm
3050
8 l/m
8 l/m
7 l/m
6 l/m
6 l/m
6 l/m
5.5 lm
4.5 l/m
3 l/m
3 l/m
2.5 l/m
2 l/m
3080
10 l/m
8 l/m
4 l/m
2 l/m
8 l/m
7 l/m
3.5 l/m
1.8 l/m
4.6 l/m
4 l/m
2.2 l/m
1.1 l/m
600 SERIES - COOLER CAPACITY DATA
Ambient 77°F/25°C
Water Vapor
Ambient 90°F/25°C
Water Vapor
Ambient 105°F/25°C
Water Vapor
12%
15%
30%
50%
12%
15%
30%
50%
12%
15%
30%
50%
620
2.5 l/m
2 l/m
1 l/m
0.6 l/m
2 l/m
1.8 l/m
0.9 l/m
0.5 l/m
1.5 l/m
1.2 l/m
0.6 l/m
0.3 lm
630
4 l/m
4 l/m
4 l/m
4 l/m
3 l/m
3 l/m
3 lm
3 l/m
2 l/m
2 l/m
2 l/m
2 l/m
640
5 l/m
4 l/m
2 l/m
1 l/m
4 l/m
3.5 l/m
1.8 l/m
0.9 l/m
3 l/m
2.5 l/m
1.3 l/m
0.7 l/m
(Flow Rate, Water Vapor % and the Ambient Temperature are the three main factors to consider when sizing a gas cooler.)