Rotary Focus Calibration
a
.
In the UCP Viewer Tab, select the Focus Feature on the Viewer Tab. Using the mouse pointer
and the manual motion keys in the UCP move the carriage manually and center red alignment
pointer over the flat part of the internal cone fixture. Use the focus tool and manually focus on top
of the flat part of the internal cone fixture.
b
.
Next, select the System Tab and press the Rotary CALIBRATE button to activate the Rotary
Calibration Window. Press the Z position SAVE button ONLY. If asked to override the current
position, accept the new position. Once the Rotary Calibration window is closed the focus
carriage will re-home.
c
.
Go back to the Viewer Tab and exit the focus feature.