Power socket for 24V DC 1A 

6. Preparing the pump for use 

 
Plug the supplied PSU into your power socket and the low voltage plug into the socket on the 
left side of the unit, switch it on by pressing the ‘ON’ button. The green LED on the side and 
the display will illuminate indicating that the unit is operational. 

 

The display will temporarily read ‘BOXER 9100’ and then the display will show the value it 
has in it’s memory. 

 

Note the direction of flow. The pump can be reversed in flow direction using the slide switch, 
so that you can  aspirate any unused reagent back into the reservoir. 
 
 
 

Page 4 

Pump direction slide switch: 
Front = Bottom side dispenswe 
Back= Bottom side aspirate 

Remote pulse socket  for connection to foot switch 
(Accessory) or pulse from a robotic processor 
Accepts Ø2.5mm mono or stereo plug 
 

On/Off switch 

Power ‘ON’ LED 

Page 5 

IMPORTANT: CHOOSE THE RANGE THAT MATCHES THE ONE YOU ARE MOST 
LIKELY TO USE IN YOUR APPLICATION. 

 

7. Choosing the correct working range 

 
There are 4 working ranges to choose from. These working ranges correspond to tube 
diameters. The decimal point is shifted corresponding with the working range. 
 
R1 

ID Ø0.5mm for < 9ml  

 

max display   9.000ml 

R2 

ID Ø1mm for <  25ml  

 

max display 25.000ml* 

R3 

ID Ø2mm for <  90ml  

 

max display 90.000ml 

R4 

ID Ø3mm for <190ml  

 

max display   190.0ml 

 
* The single µl digit can not be entered in R2 & R3 

 

Working with low volume dispenses with large diameter tube will 
reduce both precision and accuracy of dispenses. 
 
Using the U/D keys scroll to the screen  displaying “Range Enter?” 
Press Enter. 
Using the U/D keys choose the range corresponding to your tube 
diameter and press Enter. 
The display will show the last volume you worked with at that 
particular range. 
 

8. Gravimetric pump calibration procedure 

 
Pump flow performance is  subject of many factors among them the condition of the 
peristaltic tube and the viscosity of the reagent. It is therefore important that you calibrate the 
pump before it is commissioned using your reagent and at periodic intervals.  
 
Carry out the calibration procedure using a volume nearest to the volume yiou are most likely 
to dispense. You can choose any volume. 
 
1. 

Carry out 5 identical dispenses into a vessel and establish the 
gravimetric volume dispensed in each operation. 

2. 

Calculate the mean dispensed volume. 

3. 

Using the U/D keys scroll to display  saying “Calibrate Enter?” 

4. 

Press Enter  

5. 

The display shows “C” to indicate that you are in the 
calibration mode and next to it the range which you calibrate 
i.e. 1 to 4. 

6. 

Using the U/D keys correct the volume to your calculated 
mean volume * 

7. 

Press Enter and the display shows the volume you intended to dispense. The unit is 
now calibrated with a particular tube within your chosen working range. 

 
* Volumes in R2 & R3 will increment by steps of 10µl 
 
The unit will store the calibration you carried out for each individual range. 
 

Summary of Contents for BOXER 9100

Page 1: ...1 0mm complete with connectors Cat No 9000 504 Silicone Tubing IDØ2 0mm complete with connectors Cat No 9000 505 Silicone Tubing IDØ3 0mm complete with connectors Cat No 9000 510 Santoprene Tubing IDØ1 0mm complete with connectors Cat No 9000 501 Santoprene Tubing IDØ2 0mm complete with connectors Cat No 9000 502 Santoprene Tubing IDØ3 0mm complete with connectors Cat No 9000 518 Santoprene tubing...

Page 2: ...ical compatibility chart 1 A Unpacking Remove the packing materials unpack the pump controller and the power supply unit Make sure that you have got all the components as specified in Box Contains last page of this manual Please contact your supplier if you notice any one of the components is missing or damaged Note Do not attempt to assemble a unit using damaged components Retain the packaging so...

Page 3: ... pressure by lifting the pressure lever 3 C Replacing the tube Tube need replacing when worn out or wherever you intend to change the dispense volume range in which you work for smaller or larger dispense volumes Lift the pressure lever and replace the tube making sure that you have installed the tube centrally and that the tube is fully engaged by the tube guides Page 2 Pressure lever Clamp 4 The...

Page 4: ...t be entered in R2 R3 Working with low volume dispenses with large diameter tube will reduce both precision and accuracy of dispenses Using the U D keys scroll to the screen displaying Range Enter Press Enter Using the U D keys choose the range corresponding to your tube diameter and press Enter The display will show the last volume you worked with at that particular range 8 Gravimetric pump calib...

Page 5: ...ered volume in the reversed direction Page 7 13 Interval Dispense You can set the unit to dispense any volume for a number of times in predefined intervals This feature is useful in repeatable operations either in manual mode or when the dispense head is held by a robotic arm Intervals can be set between 0 5 sec to 100 minutes To set repeat dispensing 1 Using the U D keys scroll to the screen sayi...

Page 6: ...de into stored values and therefore the calculator will round up or down to the nearest whole figure 5 Q Can I leave the unit switched ON throughout the day A Yes you can The dispenser requires very little energy when the pump is not running so continuous power into the unit will not harm it 6 Q Can I dispense hot reagents using the dispenser A Yes the dispenser to works safely with media temperat...

Page 7: ...1 Santoprene tubing ID Ø2 00mm1meter running length Cat No 9000 506 Santoprene tubing ID Ø3 00mm1meter running length Cat No 9000 515 Silicone tubing ID Ø1 00mm1meter running length Cat No 9000 507 Silicone tubing ID Ø2 00mm1meter running length Cat No 9000 509 Silicone tubing ID Ø3 00mm1meter running length Cat No 9000 508 Tube clamp for continuous Ø1 0mm tube set of 2 Cat No 9000 601 Tube clamp ...

Reviews: