Meaning of Some Symbols and Troubleshooting:
10.1.
Meaning of Symbols:
0
﹉﹉﹉﹉
Zero is over the setting range
0
﹍﹍﹍﹍
Zero point is below the setting range
Ad
﹉﹉﹉
Signal to ADC is over max. range
Ad
﹍﹍﹍
Signal to ADC is below min. range
﹉﹉﹉﹉
Weight is over upper limitation, or display data is over limitation
﹍﹍﹍﹍
Weight is below lower limitation
EEP.E1
CONFIG or CAL parameters are not correctly set
EEP.E2
USER parameter is not correctly set
Lo.bAt
Battery voltage is lower than setting.
CAP.- - -
Next displaying content is Capacity
CAP.ER
Parameters about Capacity is not correct
CAL.P
x
Calibration on point(x)
CAL.OFF
Calibration Seal Switch is on OFF position
CAL.ON
Calibration Seal Switch is on ON position
CAL.Er
Calibration error, maybe input data or loaded weight is incorrect, unstable, un-linear
CAL.End
Calibration is end
OFF
Power OFF the indicator
STB.ER
Unstable time is lager than setting of USER-HOLD-STB.TIM
AcC.
xxx
Accumulation times is xxx
PR.TARE
To Preset TARE weight
COMP
To go to input COMPARE data mode
HIGH
To input HIGH limitation data of Comparison
LOW
To input LOW limitation data of Comparison
SPL.Lo
Sample load weight of low point.
SPL.HI
Sample load weight of high point.
SPL.PWT
Sample goods weight to calculate piece weight
INP.PCS
input pieces number of weighted goods
UNIT.KG
Unit kg is selected
UNIT.LB
Unit lb is selected
PWT.ER
Piece weight is error, it’s too small (<0.5d).
SPL.PCT
Sample goods weight to calculate
INP.PCT
input percentage of weighted goods
PCT.ER
Unit-Percentage -Weight is too small (<0.5d).
CACU.ER
Internal calculation overflows