3.7. Shorts Test - Master and Slave Units combined
3.7.1. From the “Shorts Test” menu, press F2 to start the shorts test.
A. “Run Shorts Test” menu is displayed. Wait for test to complete
B. If “Shorts Test Fail” is displayed, press F2 to review the faults. Follow the displayed menu
options to review each fault or return to retest.
C. When “Shorts Test Pass” is displayed, press F3 to continue to “Leakage test” menu.
3.8. Leakage Test
3.8.1.From the “Leakage Test” menu, press F2 to initiate leakage testing; allow test to finish.
A. If “Leakage Test Fail” is displayed, press F2 to review the faults. Follow the displayed
menu options to review each fault or return to retest.
B. When “Leakage Test Pass” is displayed, press F3 to continue to “Continuity Test.”
3.9. Continuity Test
3.9.1. From the “Continuity Test” menu, press F2 to start continuity testing; allow test to finish.
A. If “Continuity Test Fail” is displayed, press F2 to review the faults. Follow the displayed
menu options to review each fault or return to retest.
B. When “Continuity Test Pass” is displayed, press F3 to continue to “Test Complete” menu.
3.9.2. Press F3 to exit or F1 to continue to the Main menu.
4. Specifications
4.1. Physical
4.1.1.Size
The MU Cable Tester System in comprised of two tester units, each housed in a durable
plastic case that measures as follows:
Width: 20”
Depth: 16”
Height : 7.5”
The snap-on vinyl pouch attaches to the top of the case. The additional height of the storage
pouch is less than one inch when empty, and about 1.5 inches with the battery charger/power
adapter and chassis test lead are stored inside.
4.1.2.Weight
The weight of each MU cable tester unit with battery installed and accessory pouch with battery
charger/power adapter and chassis test lead is:
27.5 lbs.
4.2. Electrical
The following electrical parameters must not be exceeded. Operation in excess of these limits
may cause damage to the MU Cable Tester.
16814-99 Rev. A