background image

Specification 

1. UWA-N   

Capacity 

3Kg 

6Kg 

15Kg 

30Kg 

Division-A 

0.5g 

1g 

2g 

5g 

Division-B 

0.2g 

0.5g 

1g 

2g 

Division-C 

0.1g 

0.2g 

0.5g 

1g 

Mini. Cap. 

10g/4g/2g 

20g/10g/4g 

40g/20g/10g 

100g/40g/20g 

Pan size 

285 x 240 mm 

Power 

AC: 110V/220V; DC:4V 4Ah 

 

2. BSW-N/ BSW-Q 

Capacity 

60Kg 

150Kg 

300Kg 

Division-A 

10g 

20g 

100g 

Division-B 

5g 

10g 

50g 

Division-C 

2g 

5g 

---- 

Mini. Cap. 

200g/100g/400g 

400g/200g/100g 

2kg/1kg 

Pan size 

400 x 500 mm 

(size can be changed as customers’ request)

 

Power 

AC: 110V/220V; DC:4V 4Ah 

 

 

 

Preparation 

 
This product is intended for indoor use. 

  Select a suitable work area. 

  Work area should be relatively free from drafts and vibrations. 

  Work surface should be level and rigid. 

  Do not locate near magnetic materials or equipment/instruments 

which use magnets in their design. 

  Avoid areas which have variations in room temperatures or have 

excessive room temperatures. Room temperatures above 40

°

or 0

°

C could affect instrument operation and accuracy. 

 

-2- 

Calibration 

F06-USr (Quick calibration) 
When it displays   
 
Press the key   
 
It display                         

 

Put any 

STANDARD WEIGHT

 on the pan. After a little while, 

 
It will display                      , press the key            .   
 
The scale will finish calibration. 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

-7- 

F6-USr

 

P

 

0.000

 

    PUt

 

F6-USr

 

CE

 

Summary of Contents for UWA-N

Page 1: ...MODEL NO UWA N BSW N BSW Q Note 1 Please recharge the scale before first operation 2 Please warm up the scale over 5 minutes ...

Page 2: ...nstrument Disclaimer Notice Calibrate your instrument using reference weights of the appropriate tolerance class An instrument can be no more accurate than the standard to which it has been compared For assistance in the selection of reference weights please contact the factory Caution Changes or modifications not expressly approved by the manufacturer could void the user s authority to operate th...

Page 3: ...paration This product is intended for indoor use Select a suitable work area Work area should be relatively free from drafts and vibrations Work surface should be level and rigid Do not locate near magnetic materials or equipment instruments which use magnets in their design Avoid areas which have variations in room temperatures or have excessive room temperatures Room temperatures above 40 C or 0...

Page 4: ... 0 not print 1 press P to print 2 auto print when weighing 3 continuous print FA oFF Auto turn off 5 10 15 20 25 30 dis Fb ALr Hi low alarm range In our off FC nus Unit on or off U1 kg U2 g U3 lb Def U1 on U2 on U3 off 6 Function keys Numeric keys Used for setting numeric data for tare weight sample number sample weight or limit number of checking To calculate total weight Captures a new center of...

Page 5: ... Use the key to enter the OVER UNDER limit settings 2 It will display L0 0000 Use numeric key and key to set the low limitation you need 4 3 It will display H0 0000 Use numeric key and key to set the low limitation you need 4 Please be sure that hi limitation must higher than lo limitation 5 Clear the Hi Lo setting D 2 BSW N BSW Q hi lo operation 1 Use the key to enter the OVER UNDER limit setting...

Reviews: