     temperature, has a half-life of about 16 minutes in 1% of ozone solution and about 25 minutes in the air.
11.  Pure ozone, if receives impact or friction, will explode and decompose. High concentrations of ozone 

when heated is also prone to explosion. But as long as care is taken, such accidents are rather rare. 
Ozone should not have high concentrations of flammable and explosive gases at the disinfection site 
when it is used for air disinfection.

12. The user use 3M8514 welding protection mask (or equal grade protection masks) when use ozone to 

disinfect, and can process ozone protection which at 10 times occupation touch limit potency under 
1PPM

13. This product set enough safety device to guarantee the safety of users. The daily item to item check is 

the necessary condition to guarantee that safety use the equipment. Please process item to item check 
according to the requirements in the specification, need stop use the equipment immediately once 
found the problem. 

14. Because of the instability of ozone, ultraviolet light accelerates the decomposition of ozone and 

converts it back to oxygen. After the ozone work is completed, the equipment switches to UV mode to 
quickly reduce the ozone concentration in the equipment. The user must never open the cabinet doors 
before the device has completed the ozone disinfection operation (before the end of the timer 
countdown in the monitor)

15. Users are advised to equip users with ozone concentration detection instruments to regularly detect 

whether the leaked ozone value of the equipment exceeds the safe value: long term populated 
enclosed spaces, such as houses, apartments, hospitals and offices, ozone must be less than 0.1PPM in 
English. If ozone leakage value is found to have exceeded the above standard, the equipment must be 
immediately stopped.

To guarantee the disinfect effect of UVC:

1.   Whether UV light can effective disinfect depends on the amount of exposure received by the 

disinfection object, which depends on the radiation power of the UV light source, the distance of the 
disinfection object from the UV light source, and the duration of the exposure to UV light.

2.  When killing common microorganisms, the irradiated dose should reach 10,000 μW/cm²; 100,000 μ

W/cm² is the recommended dose to kill bacterial spores. The radiation dose should not be less than 
100,000 μW/cm² when the target microorganism for disinfection is unknown. 

3.  According to the radiation intensity of the UV light source of this product and the general distance of 

the UV light source from the irradiated object, it is recommended that:

•    In order to achieve the radiation dose of 10,000 μW/cm² to meet the general needs of daily use, the 

exposure time should be at least above 60 seconds

•    In order to achieve the radiation dose of 100,000 μW/cm² to meet the needs of the professional 

environment, the exposure time should be at least above 500 seconds

To guarantee the ozone disinfect effect:

1.   Ozone can kill bacterial reproductions, viruses, fungi, etc., and can destroy botulinum toxin. In order for 

ozone to be effective as a disinfectant, its concentration must be much higher than that that can be 
safely tolerated by humans and animals. Ozone has a powerful killing effect on microorganisms in the 
air, and at the same time has the odor-removal characteristics, 30minutes of 20mg/m³ (9.35PPM) 
concentration of ozone in a sealed environments can achieve disinfection effect. Ozone on the surface 
of the items of the microorganism also has a killing effect, requiring 60mg/m³ (28.04PPM),  relative 
humidity of 70%, and exposure of 60 minutes to 120 minutes, to achieve disinfection effect.

Summary of Contents for ULTIMATE UVP101R

Page 1: ...μW cm REV ODS Reverse Obstacle Detection System 2cm Sensor distance 7 4V 2600 mAh Li ion UV 3 3 5h O3 15 16h Input Output 5V 2A Work light specifications Center box Irradiance 240μW cm Intelligent UV OZONE Disinfection Box ULTIMATE UVP101R UVC Light germicidal system intends for use in general purpose ...

Page 2: ...ill start automatically If detecting the distance between the light source of disinfection light head and the surface of articles to be disinfected exceeds 2cm it will automatically stop the disinfection work and reach the safe energy efficient effects In case of disinfecting the surface of articles by independently hand held using it is usually able to kill 99 9 bacteria and virus on the surface ...

Page 3: ...tton during the disinfection will not be effective 10 It will automatically stop the ozone work in the last 10 minutes of preset time in the course of ozone disinfection and activate the ultraviolet automatically to perform 10 minute ozonolysis 11 If the battery level is lower than 20 and the battery level displays in red it is required to charge as soon as possible to avoid that the remaining bat...

Page 4: ...one disinfection As for the ordinary article disinfection it is suggested to disinfect for 60 minutes As for the professional article disinfection it is suggested to appropriately extend the disinfection time SAFETY WARNING 1 Please read and understand this user manual before use to ensure that the product is safely and properly used 2 Our Company shall not be liable for product damage and other i...

Page 5: ...ing to the FDA Although ozone has been reported to have adverse physiological effects on the central nervous system heart and vision its main physiological effect is to stimulate the mucous membranes Inhalation of ozone causes enough irritation to the lungs leading to pulmonary edema The onset of pulmonary edema is usually delayed by several hours after exposure As the hazards of ozone are widely ...

Page 6: ...eakage value is found to have exceeded the above standard the equipment must be immediately stopped To guarantee the disinfect effect of UVC 1 Whether UV light can effective disinfect depends on the amount of exposure received by the disinfection object which depends on the radiation power of the UV light source the distance of the disinfection object from the UV light source and the duration of t...

Page 7: ...es at a time using ozone concentrations of 35PPM 4 After 10 tests of six different leather material colors samples the results show color fading and hardening of the material especially the brighter samples whose fading was more pronounced 5 The same conditions were tested on stainless steel aluminum gold plated PIN needles silicone cotton pads cotton and ABS plastics and no significant decomposit...

Page 8: ...structions WARNING UV C emitted from this product Avoid eye and skin exposure to unshielded product Follow installation instructions and user manual www uvpluss com AEC Lighting Solutions Co Ltd Add 2548 Bao an Highway Jiading District Shanghai 201801 China Tel 400 083 8757 www aeclight reel com ...

Reviews: