(Start Delay) for user to leave the room

• After clicking the option to be set, a numeric input keyboard 
  will pop up:
• According to actual needs, after setting the desired parameters,
  the total working time (Sterilization Time) and the total amount 
  of disinfectant liquid (Total Dose) will be automatically calculated. 
  According to the total amount of disinfectant liquid, the user needs 
  to pour the corresponding amount into the disinfectant bottle. After adding the disinfectant, and after 

setting the Start Delay to leave the room, click the Start button, and ensure that all personnel leaves 

  the room quickly, and the disinfection operation will start immediately after the countdown is over. 
• When equipment is in delay countdown or already in disinfection operation, the use can press “Stop” 

(“Start” button becomes “Stop” when equipment has started countdown) to immediately stop the 
equipment operation and return to pre start status.

7. Disinfect operation process

There are a total of 3 stages in the operating cycle of the equipment.
A: First, spray sterilization (automatically calculated according to room area)

Move the equipment to the sterilizing area, place on selected location, and plug in the power 
supply. Turn on the equipment power switch, and enter the settings interface, set the room size and 
delay time settings according to the actual sterilization area size and equipment location. Enter 
actual volume of the sterilization area into the Room Volume; Delay time can be set according to 
the user’s needs. After each setting is set, the interface will automatically display the estimated 
dosage, total added amount, sterilization time and running time.
Move the equipment in place, then open any inner doors of each room within the area to be 
sterilized to ensure clear air flow, and then close the outermost door. Confirm the initial 
temperature and humidity in the room to ensure that it is within the normal range, the temperature 
is 18-26 degrees Celsius, and the humidity is 30-60%. Turn off the air conditioning system, and 
after all personnel have left the sterilization area, click the [start] button to turn on the equipment 
and enter the sterilization process. Note that one full bottle is 1000ml, so if the user sets the room 
volume above 200m³ as per below picture A, the sprayer will stop after the bottle is depleted. The 
UI display will then display per below picture B (total dose reduced since 1 full bottle is already 
used up and run time shows one spray session has concluded):

The user has to then refill the bottle with disinfectant, press continue and then vacate the room, the 

equipment will resume after the countdown is finished. This cycle (conclude spraying, refill bottle, 
resume spraying) continues until the total dose is 0. Then the equipment will move on to the next 
phase (dwell period)

A

B

Summary of Contents for UVP801C

Page 1: ...roxide Disinfector UVP801C 484x270x314mm 7 1kg 80m s 500m Log6 Work light specifications 16ml min Sterilization effect Sterilization volume Sterilization speed 10 µm Hydrogen peroxide solution 22000rps Max 1200W Output 1 ...

Page 2: ...est condition Laser particle size analyzer One hour test 18 points space disinfection test Thank you for using our product our company will provide you with the best service and our Dry fog hydrogen peroxide disinfector is easy to use safe and reliable This manual contains detailed information on how to use the sprayer equipment management personnel and operators must fully read this manual prior ...

Page 3: ...ea parameters memory function Built in calculation software Sturdy structure easy maintenance One single equipment can disinfect 20 500m3 space The whole process takes 2 4 hours greatly shortening disinfection duration Replaced methanol and can achieve Log6 sterilization rate GMR FDA and EP recommend sterilization method in cleaning area 4 Product characteristics 5 Equipment structure 1 Disinfecta...

Page 4: ... the event of malfunction or breakdown grounding provides a path of least resistance for electric current to reduce the risk of electric shock This appliance is equipped with a cord having an equipment grounding conductor and a grounding plug The plug must be plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes and ordinances DANGER improper...

Page 5: ... the like extending from the adaptor must be connected to a permanent ground such as a properly grounded outlet box cover Whenever the adaptor is used it must be held in place by a metal screw Turning on the power switch will enter into below user interface 6 Parameters setting User is required to set the room size the residual dose volume of the disinfectant then click Start Start Delay countdown...

Page 6: ...me settings according to the actual sterilization area size and equipment location Enter actual volume of the sterilization area into the Room Volume Delay time can be set according to the user s needs After each setting is set the interface will automatically display the estimated dosage total added amount sterilization time and running time Move the equipment in place then open any inner doors o...

Page 7: ...to achieve the desired sterilization effect How much additional time is directly proportional to the reduced solution percentage ie if 3 solution is used the equipment must be set to spray twice the needed amount which also means twice the amount of time compared to 6 solution Dwell time and residue removal duration is not affected User safety The user use 3M8514 welding protection mask or equal g...

Page 8: ...e mask or similar grade protective mask When repairing the equipment in addition to wearing the above protective mask the equipment should be kept in a well ventilated environment Drop off the residual disinfectant after each time use finished use the purified water to atomizing trial wash pipeline and spraying head under the allowed condition for maintenance purpose 9 Daily maintenance Do not sta...

Reviews: