background image

1. INSTALLATION 

Because of metrological legislation, installation/some 

metrological parameter settings are limited to be done 

by authorized personnel only. Do not attempt to change 

any of the built-in parameters. Contact your dealer for 

installation and technical assistance. 

 

CAUTION: 

This unit is legal for trade only when it is sealed (and/or 

stamped) and bearing a serial number. Do not attempt 

to break the seal (or stamp) affixed to this indicator 

or remove the serial number. Contact your dealer for 

more information and after sales service. 

 

For most accurate weighing result, do not use the 

indicator in where or when the environment condition 

falls beyond as those listed on 

SPECIFICATIONS

 

Do not attempt to open this unit or conduct any trouble 

shootings other than those listed on 

TROUBLE SHOOTING

 

 

 

 

2. SPECIFICATIONS 

 

Model No. 

ABM/AFM/AFS/ATM INDICATOR 

Max. 

USER DEFINED 

n

max

 

3000 

(NOTE 1)

 

Tare Range 

1/3 Max. ¡ e 

(NOTE 2)

 

Power-on Zero Range 

¡Ó10% Max. 

Zero Range 

¡Ó2% Max. 

Min. Load 

20e 

(NOTE 3)

 

Max. measuring range 

9.0mV 

Min. signal voltage per 
verification scale interval 

1.5  V 

 

Power voltage requirements 

6.0 VDC (BY EXTERNAL POWER 
ADAPTOR OR RECHARGEABLE 
BATTERY) 

Minimum battery voltage 

5.5 VDC 

Load cell Excitation voltage 

5 VDC 

Minimal load cell impedance 

85  

Maximal load cell impedance 

1000  

Load cell connection 

4 wires 

Operation 
Environment 

0

o

~40

o

Non-condensed. R.H.

¡

85% 

Specifications subject to change without notice. 

 

NOTE 1: 

The recommend division when application is legal for 

trade  (OIML mode) is 3000 and 6000 to 7500  when 

application is not legal for trade (normal mode). The 

maximum display division under OIML mode when Extended 

Display function engaged is 30000, the same as under 

non-OIML mode. 

Summary of Contents for ABM

Page 1: ...FULLY BEFORE ATTEMPT TO OPERATE THIS INDICATOR JUNE 2003 REV 2 Specifications subject to change without prior notice CONTENTS 1 INSTALLATION 2 SPECIFICATIONS 3 KEYBOARD LAYOUT AND DESCRIPTION 4 INITIAL SETUP 5 INSTRUCTION FOR USE 6 TROUBLE SHOOTING 7 DAILY CARE AND MAINTENANCE APPENDIX APPEARANCE ...

Page 2: ... shootings other than those listed on TROUBLE SHOOTING 2 SPECIFICATIONS Model No ABM AFM AFS ATM INDICATOR Max USER DEFINED nmax 3000 NOTE 1 Tare Range 1 3 Max e NOTE 2 Power on Zero Range Ó10 Max Zero Range Ó2 Max Min Load 20e NOTE 3 Max measuring range 9 0mV Min signal voltage per verification scale interval 1 5 V Power voltage requirements 6 0 VDC BY EXTERNAL POWER ADAPTOR OR RECHARGEABLE BATTE...

Page 3: ...l tare range NOTE 3 For Multi range or interval change e to e1 3 KEYBOARD LAYOUT AND DESCRIPTION MODEL SHOWN ATM 1 ON OFF KEY Press this key to turn indicator on and off 2 MODE KEY Press this key to shift among different weight units and or turn backlight on off 3 M KEY Press this key to accumulate current weight to memory ...

Page 4: ...dicates the metric unit is being employed lb indicates the avoirdupois unit is being employed 10 IN CHARGE INDICATOR Red color Recharging on process Green color Charging completed 11 AVERAGE INDICATOR This indicator flashes to indicate the digital motion filtering function is being employed 12 M INDICATOR This indicator appears to indicate memory contains of stored data 13 MR INDICATOR This indica...

Page 5: ...on The scale will power off automatically if it is not used for 4 minutes Follow the below steps to enable disable Auto Power Off Function a Scale is off b Press and hold TARE then press ON OFF c Scale displays F 1 d Press TARE three times to F 4 e Press MODE to shift between 0 _OFF or 4 _OFF To enable Auto Power Off Function press TARE when 4 _OFF appears To disable Auto Power Off Function press ...

Page 6: ...RE when display shows Filt 1 To disable Animal Weighing Function press TARE when display shows Filt 0 f After TARE is pressed indicator display F 7 g Press MODE to enter next set up or ZERO to restart the scale 4 2 5 Select Serial Output Mode There are two selectable transmission formats to choose according to end user preference For which FORMAT 1 is continuous transmission when weight is stabili...

Page 7: ...r TARE is pressed scale display F 9 g Press MODE to enter next set up or ZERO to restart the scale 4 2 7 Select Auto Tare Function This scale is equipped with auto tare function which tares off the initial weight automatically when this function is engaged The tare weight is automatically cleared when everything is removed from platter a Scale is off b Press and hold TARE then press ON OFF c Scale...

Page 8: ...f would be employed when the scale is turned on again NOTE To comply with the law of certain countries and certain approval requirements the avoirdupois weight units may be disabled Contact your dealer for more information 5 4 TURN BACKLIGHT OPTIONAL ON OFF Follow the below steps to turn on and off backlight if purchased a Scale is on b Press and hold ZERO until backlight is activated or pressMODE...

Page 9: ... pressed the NET INDICATOR would appear Refer to 5 2 for weighing procedures Weight displayed under is the net weight of the subject matter Refer to SPECIFICATIONS for maximum tare range NOTE When scale is set at OIML and application is legal for trade User only needs to press TARE once when tare off a weight The tare function will take place when reading is stabled When scale is set at Normal and...

Page 10: ...mulated follow by a value The value is the total accumulated weight stored in memory When the total accumulated weight is being displayed MR INDICATOR appears to indicator that the value being displayed is the total accumulated weight Indicator will return to normal display status after 3 seconds 5 8 3 Memory Clear All transactions stored in memory by pressing MR and TARE at the same time CAUTION ...

Page 11: ... or lb Check Is the load applied to scale excess the rated capacity Action Remove all loads and try again Syndrome Indicator displays Check Is the load cell signal cables connected to the indicator properly Action Recheck signal cable connections Check Is load cell working properly Action Check load cell input output Syndrome Display shows 00000 after counting down Check Is the load applied to pla...

Page 12: ...etone volatile solvent thinner or alcohol for cleaning 7 3 Verify the accuracy of indicator periodically Re calibrate the indicator if necessary NOTE In some countries calibration requires authorized qualified agent Contact your dealer for more information 7 4 Store indicator scale in a dry and clean place APPENDIX APPEARANCE MODEL SHOWN ABM AFM MODEL SHOWN AFS ...

Page 13: ...irements of the following EC directives correspond au modele decrit dans le certificat d approbation CE de type aux exigences de la directive 90 384 CEE modifiee et aux exigences des directives CE suivantes Diese Erklarung gilt nur in Verbindung mit einer Konformitatsbescheinigung einer benanten Stelle This Declaration of conformity is only valid with a certificate of conformity issued by a notifi...