4. INITIAL SETUP 

4.1 PLACING THE SCALE 

In order to obtain an accurate weighing result, the 

complete weighing instrument (

hereinafter referred as 

the scale

) must be placed on a strong and level surface. 

 

4.2 INTERNAL SETTING 

4.2.1 Display Segment Check 

User can enter this function to check to see if the display 

and backlight (if purchased) are functioning properly. 

a.  Scale is off 

b.  Press and hold 

TARE

, then press 

ON/OFF

 

c.  Scale displays 

F.1

 

d.  Press 

TARE

 once to 

F.2

 

e.  Press 

MODE

 and all segments will light up to allow 

user to check the condition of display 

f.  Press any key to exit and 

TARE

 for next function 

setup or 

ZERO

 to restart the scale 

 

4.2.2 Select Auto Power Off Setting 

The scale is equipped with 

Auto Power Off Function

. The 

scale will power off automatically if it is not used 

for 4 minutes. Follow the below steps to enable/disable 

Auto Power Off Function.

 

a.  Scale is off 

b.  Press and hold 

TARE

, then press 

ON/OFF

 

c.  Scale displays 

F.1

 

d.  Press 

TARE

 three times to 

F.4

 

e.  Press 

MODE

 to shift between 

0._OFF

 or 

4._OFF

 

-To enable 

Auto Power Off Function

, press 

TARE

 when 

4._OFF

 appears 

-To disable 

Auto Power Off Function

, press 

TARE

 when 

0._OFF

 appears 

f.  Press 

ZERO

 to restart the scale or 

TARE

 for next 

function setup 

 

4.2.3 Select RS-232 Baud Rate & Protocol 

This scale is equipped with 3 RS232C output baud rate 

and 2 transmission 

protocol settings for user to select.  

a.  Scale is off 

b.  Press and hold 

TARE

, then press 

ON/OFF

 

c.  Scale displays 

F.1

 

d.  Press 

TARE

 four times to 

F.5

 

e.  Press 

MODE

 to enter 

f.  Press 

M+

 to select baud rate of 4800, 9600 or 19200 

g.  Press 

MODE

 again to enter transmission protocol 

setup 

h.  Press

 M+

 to select protocol of P=N81 or P=E71  

Summary of Contents for ABM

Page 1: ...FULLY BEFORE ATTEMPT TO OPERATE THIS INDICATOR JUNE 2003 REV 2 Specifications subject to change without prior notice CONTENTS 1 INSTALLATION 2 SPECIFICATIONS 3 KEYBOARD LAYOUT AND DESCRIPTION 4 INITIAL SETUP 5 INSTRUCTION FOR USE 6 TROUBLE SHOOTING 7 DAILY CARE AND MAINTENANCE APPENDIX APPEARANCE ...

Page 2: ... shootings other than those listed on TROUBLE SHOOTING 2 SPECIFICATIONS Model No ABM AFM AFS ATM INDICATOR Max USER DEFINED nmax 3000 NOTE 1 Tare Range 1 3 Max e NOTE 2 Power on Zero Range Ó10 Max Zero Range Ó2 Max Min Load 20e NOTE 3 Max measuring range 9 0mV Min signal voltage per verification scale interval 1 5 V Power voltage requirements 6 0 VDC BY EXTERNAL POWER ADAPTOR OR RECHARGEABLE BATTE...

Page 3: ...l tare range NOTE 3 For Multi range or interval change e to e1 3 KEYBOARD LAYOUT AND DESCRIPTION MODEL SHOWN ATM 1 ON OFF KEY Press this key to turn indicator on and off 2 MODE KEY Press this key to shift among different weight units and or turn backlight on off 3 M KEY Press this key to accumulate current weight to memory ...

Page 4: ...dicates the metric unit is being employed lb indicates the avoirdupois unit is being employed 10 IN CHARGE INDICATOR Red color Recharging on process Green color Charging completed 11 AVERAGE INDICATOR This indicator flashes to indicate the digital motion filtering function is being employed 12 M INDICATOR This indicator appears to indicate memory contains of stored data 13 MR INDICATOR This indica...

Page 5: ...on The scale will power off automatically if it is not used for 4 minutes Follow the below steps to enable disable Auto Power Off Function a Scale is off b Press and hold TARE then press ON OFF c Scale displays F 1 d Press TARE three times to F 4 e Press MODE to shift between 0 _OFF or 4 _OFF To enable Auto Power Off Function press TARE when 4 _OFF appears To disable Auto Power Off Function press ...

Page 6: ...RE when display shows Filt 1 To disable Animal Weighing Function press TARE when display shows Filt 0 f After TARE is pressed indicator display F 7 g Press MODE to enter next set up or ZERO to restart the scale 4 2 5 Select Serial Output Mode There are two selectable transmission formats to choose according to end user preference For which FORMAT 1 is continuous transmission when weight is stabili...

Page 7: ...r TARE is pressed scale display F 9 g Press MODE to enter next set up or ZERO to restart the scale 4 2 7 Select Auto Tare Function This scale is equipped with auto tare function which tares off the initial weight automatically when this function is engaged The tare weight is automatically cleared when everything is removed from platter a Scale is off b Press and hold TARE then press ON OFF c Scale...

Page 8: ...f would be employed when the scale is turned on again NOTE To comply with the law of certain countries and certain approval requirements the avoirdupois weight units may be disabled Contact your dealer for more information 5 4 TURN BACKLIGHT OPTIONAL ON OFF Follow the below steps to turn on and off backlight if purchased a Scale is on b Press and hold ZERO until backlight is activated or pressMODE...

Page 9: ... pressed the NET INDICATOR would appear Refer to 5 2 for weighing procedures Weight displayed under is the net weight of the subject matter Refer to SPECIFICATIONS for maximum tare range NOTE When scale is set at OIML and application is legal for trade User only needs to press TARE once when tare off a weight The tare function will take place when reading is stabled When scale is set at Normal and...

Page 10: ...mulated follow by a value The value is the total accumulated weight stored in memory When the total accumulated weight is being displayed MR INDICATOR appears to indicator that the value being displayed is the total accumulated weight Indicator will return to normal display status after 3 seconds 5 8 3 Memory Clear All transactions stored in memory by pressing MR and TARE at the same time CAUTION ...

Page 11: ... or lb Check Is the load applied to scale excess the rated capacity Action Remove all loads and try again Syndrome Indicator displays Check Is the load cell signal cables connected to the indicator properly Action Recheck signal cable connections Check Is load cell working properly Action Check load cell input output Syndrome Display shows 00000 after counting down Check Is the load applied to pla...

Page 12: ...etone volatile solvent thinner or alcohol for cleaning 7 3 Verify the accuracy of indicator periodically Re calibrate the indicator if necessary NOTE In some countries calibration requires authorized qualified agent Contact your dealer for more information 7 4 Store indicator scale in a dry and clean place APPENDIX APPEARANCE MODEL SHOWN ABM AFM MODEL SHOWN AFS ...

Page 13: ...irements of the following EC directives correspond au modele decrit dans le certificat d approbation CE de type aux exigences de la directive 90 384 CEE modifiee et aux exigences des directives CE suivantes Diese Erklarung gilt nur in Verbindung mit einer Konformitatsbescheinigung einer benanten Stelle This Declaration of conformity is only valid with a certificate of conformity issued by a notifi...

Reviews: